{{Spec | Temperature  | –62°C to +150°C (±2°C cal. probe, ±4°C uncal. probe), +150°C to +240°C (+0/-6°C or +2/-8°C any probe, uncal.) }}
{{Spec | Temperature  | –62°C to +150°C (±2°C cal. probe, ±4°C uncal. probe), +150°C to +240°C (+0/-6°C or +2/-8°C any probe, uncal.) }}
{{EndSpecs}}
{{EndSpecs}}
==Rear interface==
The DM501A has a rear input, selectable by a front-panel button, on B28 (Hi) / A29 (Lo).  Unlike the DM501 (non-A), it has no digital outputs.
Input voltage on the rear interface is limited to 200 V peak / 100 V RMS.


==Pictures==
==Pictures==
<gallery>
<gallery>
Tek dm501a on.jpg|DM501A
Tek dm501a on.jpg|DM501A

Revision as of 00:36, 14 September 2023

Tektronix DM501A
4½ digit True RMS multimeter
Tektronix DM501A

Produced from 1980 to 1984

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix DM501A is a 4½ digit True RMS multimeter plug-in for the TM500 system.

It replaced the DM501 (which suffered from accuracy limitations) in 1980. It also has a temperature input, requiring the P6601 PT-100 probe with a 2-pin LEMO S-series connector.

Key Specifications

DC Voltage 200 mVDC to 200 VDC in decade ranges (0.05% + 0.01% (200 mV: 0.015%) FS), 1000 VDC (0.05% + 0.02% FS); Input resistance 10 MΩ, CMRR >100 dB at DC, 80 dB at 50 and 60 Hz (1 kΩ unbal.), NMRR 60 dB at 50 or 60 Hz
AC Voltage 200 mVDC to 200 VAC in decade ranges, and 500 VAC (0.6% + 0.05% (500 V:0.2%) FS, True RMS); Input impedance 10 MΩ // ≤160 pF, CMRR 60 dB at 50 or 60 Hz (1 kΩ unbal.)
DC Current 200 μA to 2 A in decade ranges, 0.2% + 0.015% FS
AC Current 200 μA to 2 A in decade ranges, 0.6% + 0.05% FS (for 20 Hz to 10 kHz sine >5% FS)
Ohms 200 Ω to 20 MΩ in decade range; current 1 mA, 1 mA, 100 μA – 0.1 μA respectively; 0.15% (20 MΩ: 0.5%) + 0.015% FS
Temperature –62°C to +150°C (±2°C cal. probe, ±4°C uncal. probe), +150°C to +240°C (+0/-6°C or +2/-8°C any probe, uncal.)

Rear interface

The DM501A has a rear input, selectable by a front-panel button, on B28 (Hi) / A29 (Lo). Unlike the DM501 (non-A), it has no digital outputs. Input voltage on the rear interface is limited to 200 V peak / 100 V RMS.

Pictures

Parts

Some Parts Used in the DM501A

Part Part Number(s) Class Description Used in
AD536 156-1259-00 Monolithic integrated circuit RMS/DC converter DM501A DM502A DM5010 Keithley 177 Keithley 179 Keithley 195A
Siliconix LD120 156-1306-00 Monolithic integrated circuit 4½ digit integrating ADC (analog section) DM501A DM5010
Siliconix LD121A 156-1305-00 Monolithic integrated circuit 4½ digit integrating ADC (digital section) 7A13 DM501A