The '''Tektronix P6135(A)''' is a differential probe pair matched to minimize differences in impedance and signal delay,  
The '''Tektronix P6135(A)''' is a differential probe pair matched to minimize differences in impedance and signal delay,  
to maintain CMRR up to 20,000:1 when used with differential amplifiers such as the [[7A13]], [[7A22]] or [[11A33]].
to maintain CMRR up to 20,000:1 when used with differential amplifiers such as the [[7A13]], [[7A22]] or [[11A33]].
Note that, unlike most x10 probes, the P6135 and P6135A have an input resistance of 1 MΩ rather than 10 MΩ.


The P6135 is a [[Subminiature modular probes|subminiature modular probe]] with a limited compensation range around 15 pF.  
The P6135 is a [[Subminiature modular probes|subminiature modular probe]] with a limited compensation range around 15 pF.  
Line 20: Line 21:


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Scope Impedance Compensation Range | P6135: 1 MΩ // 13-17 pF; P6135A: 1 MΩ // 13-17 pF (red tip), 1 MΩ // 18-22 pF (black tip) and 1 MΩ // 45-49 pF (blue tip) }}
{{Spec | P6135 Compensation Range | 1 MΩ // 13-17 pF }}
{{Spec | P6135A Compensation Range | with Red Tip: 1 MΩ // 13-17 pF }}
{{Spec |                          | with Black Tip: 1 MΩ // 18-22 pF }}
{{Spec |                          | with Blue Tip: 1 MΩ // 45-49 pF }}
{{Spec | P6135 Input Impedance | ''Please Add'' }}
{{Spec | P6135A Input Impedance | with Red Tip: 1 MΩ // 10.5 ± 0.6 pF }}
{{Spec |                        | with Black Tip: 1 MΩ // 11.0 ± 0.6 pF }}
{{Spec |                        | with Blue Tip: 1  MΩ // 13.7 ± 0.6 pF }}
{{Spec | Attenuation | Adjustable to ×10 }}
{{Spec | Attenuation | Adjustable to ×10 }}
{{Spec | CMRR | 10,000:1 on [[11A33]], 20,000:1 on [[7A13]] }}
{{Spec | CMRR | 10,000:1 on [[11A33]], 20,000:1 on [[7A13]] }}
{{Spec | Bandwidth | 150 MHz}}
{{Spec | Bandwidth | 150 MHz on [[11A33]], 90 MHz on [[7A13]] }}
{{Spec | Max. input | ±500 V Maximum Common Mode Voltage }}
{{Spec | Max. Input | ±500 V (DC + peak AC) derated above 1.3 MHz }}
{{EndSpecs}}
{{EndSpecs}}



Revision as of 17:36, 22 August 2024

Tektronix P6135
150 MHz, ×10 differential probe pair
Tektronix P6135

Compatible with 1 MΩ scope inputs

Produced from 1987 to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6135(A) is a differential probe pair matched to minimize differences in impedance and signal delay, to maintain CMRR up to 20,000:1 when used with differential amplifiers such as the 7A13, 7A22 or 11A33. Note that, unlike most x10 probes, the P6135 and P6135A have an input resistance of 1 MΩ rather than 10 MΩ.

The P6135 is a subminiature modular probe with a limited compensation range around 15 pF. The P6135A is a larger compact modular probe and comes with three different tips for increased compensation range, matching the 11A33 − red, 7A13 − black, and 7A22 − blue. The P6135A model can therefore be seen as a replacement of the P6055.

Key Specifications

P6135 Compensation Range 1 MΩ // 13-17 pF
P6135A Compensation Range with Red Tip: 1 MΩ // 13-17 pF
with Black Tip: 1 MΩ // 18-22 pF
with Blue Tip: 1 MΩ // 45-49 pF
P6135 Input Impedance Please Add
P6135A Input Impedance with Red Tip: 1 MΩ // 10.5 ± 0.6 pF
with Black Tip: 1 MΩ // 11.0 ± 0.6 pF
with Blue Tip: 1  MΩ // 13.7 ± 0.6 pF
Attenuation Adjustable to ×10
CMRR 10,000:1 on 11A33, 20,000:1 on 7A13
Bandwidth 150 MHz on 11A33, 90 MHz on 7A13
Max. Input ±500 V (DC + peak AC) derated above 1.3 MHz