PG509: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 10: | Line 10: | ||
* [[Media:070-7755-00.pdf|Tektronix PG509 Manual]] | * [[Media:070-7755-00.pdf|Tektronix PG509 Manual]] | ||
}} | }} | ||
It | It is intended for use as a stimulus for making measurements and produces a fixed-amplitude rectangular positive pulse waveform. | ||
{{BeginSpecs}} | {{BeginSpecs}} | ||
Line 30: | Line 27: | ||
{{Documents|Link=PG509}} | {{Documents|Link=PG509}} | ||
{{PatentLinks|PG509}} | {{PatentLinks|PG509}} | ||
==Internals== | |||
The PG509 uses an avalanche transistor and charge line to generate the fast-rising output pulse. A voltage doubler fed by two series-connected 25 V mainframe windings provides +150 V for the avalanche stage. | |||
==Rear Interface== | ==Rear Interface== |
Revision as of 02:34, 3 November 2024
The Tektronix PG509 is a 350 ps rise-time pulse generator plug-in for the TM500 system.
It is intended for use as a stimulus for making measurements and produces a fixed-amplitude rectangular positive pulse waveform.
Key Specifications
Amplitude | 25 V (specified into 50 Ω), adjustable ±5% |
---|---|
Rise time | 350 ±50 ps |
Pulse width | 44 ±2 ns |
Repetition rate | 50 kHz ±10% |
Pretrigger out | 220 mV ±20% into 50 Ω, 115 ns ±25% before output |
Inputs |
Links
Documents Referencing PG509
- (no results)
Internals
The PG509 uses an avalanche transistor and charge line to generate the fast-rising output pulse. A voltage doubler fed by two series-connected 25 V mainframe windings provides +150 V for the avalanche stage.
Rear Interface
- A15 − Amplitude control
- B15 − Pulse inhibit
Pictures
-
-
-
-
schematic
Components
Some Parts Used in the PG509
- (no results)