P6057
The Tektronix P6057 is a 1.4 GHz ×100 probe for 50 Ω-input scopes, released along with the 7A19 amplifier in 1972.
It has an Identify switch and readout contact.
Its sister model P6056 has ×10 attenuation.
Key Specifications
Bandwidth | DC to 1.4 GHz |
---|---|
Rise time | 250 ps |
Attenuation | 100× |
Input impedance | 5 kΩ // approx. 1 pF |
Scope Input | 50 Ω |
Features | Readout Pin, ID Button |
Manual
WARNING:
This high performance device has a fragile resistor assembly in the
probe head, subject to failure in rough usage. Use with slip-on tip
whenever possible. One replacement part is provided with the package.
Contact nearest Field Office for additional parts (P/N 307-0330-00), if
required. For replacement and Calibration Procedure, refer to Pg. 3 & 4
of manual.