7S12: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(Added link to local copy of instruction manual)
m (→‎top: clean up, replaced: {{Plugin Sidebar 2 → {{Plugin Sidebar , series=7000-series scopes| → series=7000-series scopes|, title=Tektronix → manufacturer=Tektronix | type=)
Line 1: Line 1:
{{Plugin Sidebar 2|
{{Plugin Sidebar |
title=Tektronix 7S12|
manufacturer=Tektronix | type=7S12|
summary=Time Domain Reflectometer / Sampler|
summary=Time Domain Reflectometer / Sampler|
image=7s12-s6-s52-f.jpg |
image=7s12-s6-s52-f.jpg |
Line 6: Line 6:
introduced=1971 |
introduced=1971 |
discontinued=1989 |
discontinued=1989 |
series=[[7000-series scopes]]|
series=7000-series scopes|
manuals=
manuals=
* [[Media:070-1244-00.pdf | Tektronix 7S12 Instruction Manual S/N B020000-up revised June 1986]] (OCR, PDF)
* [[Media:070-1244-00.pdf | Tektronix 7S12 Instruction Manual S/N B020000-up revised June 1986]] (OCR, PDF)

Revision as of 04:04, 9 August 2021

Tektronix 7S12
Time Domain Reflectometer / Sampler
Tektronix 7S12 TDR/Sampler with S-6 and S-52 heads

Compatible with 7000-series scopes

Produced from 1971 to 1989

Manuals
Manuals – Specifications – Links – Pictures


The Tektronix 7S12 is a Time Domain Reflectometer / Sampler plug-in for 7000-series scopes. It occupies two mainframe plug-in bays (one vertical, one horizontal).

The 7S12 has two bays for S-series plug-in modules ("heads"). The right bay accepts a pulse generator or a trigger recognizer plug-in, whereas the left bay accepts a sampling plug-in.

The typical configuration is to have an S-6 on the left and an S-52 on the right for time domain reflectometry, or a sampling head like the S-4 on the left and an S-53 on the right as a generic sampler.

The 7S12 can also control an optional 7S11 through its left side contacts, adding a second input channel.

Y axis can be scaled in voltage (mV) or distance (ρ) units, X axis in time or distance.

Key Specifications

Sweep modes Single / Repetitive / Manual / External input
With S-6 and S-52
  • Pulse: 200 mV into 50 Ω
  • Rise time: < 35 ps for incident step, < 45 ps for displayed reflection
  • Jitter: < 10 ps
With S-5 and S-54
  • Pulse: > 400 mV into 50 Ω
  • Rise time: < 1.5 ns for displayed reflection of a shorted line
  • Jitter: < 20 ps
Sweep modes
  • free-running
  • single
  • manual
  • external input
Air Dielectric Range
  • 15 m (49 ft)
  • 150 m (490 ft)
  • 1500 m (4900 ft)
Weight 2.13 kg (4.7 lbs)

Links

Pictures

Inside

Measurements

TDR

Probe rise times

As a generic sampler