The '''Tektronix P6040/CT-1''' is a 35 kHz – 1 GHz current probe intended for sampling scopes.
The '''Tektronix P6040/CT-1''' is a 35 kHz – 1 GHz current probe intended for sampling scopes.


The P6040 is an 18", 50 Ω cable with a [[GR-874 connector]] for the scope end (e.g. [[S-1]], [[3S6]]) and an SMB connector on the other end to connect to one of possibly several [[CT-1]] transformers installed in a circuit to be tested.
* '''015-0041-00''' is the part number for the CT-1/P6040 combination
* '''010-0133-00''' is the part number for the P6040 alone
* '''015-0040-00''' is the part number for the CT-1 alone


{{BeginSpecs}}
The P6040 is an 18", 50 Ω cable with a [[GR-874 connector]] for the scope end (e.g. [[S-1]], [[3S6]]) and an [[SMB]] connector on the other end to connect to one of possibly several [[CT-1]] transformers installed in a circuit to be tested.
 
{{BeginSpecs}} for the CT-1
{{Spec | Sensitivity | 5 mV/mA into 50 Ω }}
{{Spec | Sensitivity | 5 mV/mA into 50 Ω }}
{{Spec | Rise time | <350 ps }}
{{Spec | Rise time | <350 ps }}
{{Spec | Frequency response | 35 kHz to 1 GHz (-30%) }}
{{Spec | Frequency response | 35 kHz to 1 GHz (-30%) }}
{{Spec | Max. current | 500 mA RMS, 100 A peak }}
{{Spec | Max. current | 500 mA<sub>RMS</sub>, 100 A peak }}
{{Spec | Input Aperture | ≥0.065" (1.6 mm) }}
{{Spec | Voltage Rating | 1000 V }}
{{Spec | Insertion Impedance | Terminated: 1 Ω // 5 μH, Un-terminated: 2 Ω // 5 μH }}
{{Spec | Capacitive Loading | 0.5 to 1.5 pF }}
{{EndSpecs}}
{{EndSpecs}}


Line 37: Line 45:
Tek_CT-1_4.jpg
Tek_CT-1_4.jpg
P6040-CT1.jpg
P6040-CT1.jpg
Tek 015-040 1.JPG | Early CT-1
Tek 015-040 2.JPG | Early CT-1
</gallery>
</gallery>


[[Category:Oscilloscope current probes]]
[[Category:Oscilloscope current probes]]

Latest revision as of 21:02, 29 November 2023

Tektronix P6040
AC current probe
P6040/CT-1 current probe

Compatible with 50 Ω scope inputs

Produced from 1963 to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6040/CT-1 is a 35 kHz – 1 GHz current probe intended for sampling scopes.

  • 015-0041-00 is the part number for the CT-1/P6040 combination
  • 010-0133-00 is the part number for the P6040 alone
  • 015-0040-00 is the part number for the CT-1 alone

The P6040 is an 18", 50 Ω cable with a GR-874 connector for the scope end (e.g. S-1, 3S6) and an SMB connector on the other end to connect to one of possibly several CT-1 transformers installed in a circuit to be tested.

Key Specifications

Sensitivity 5 mV/mA into 50 Ω
Rise time <350 ps
Frequency response 35 kHz to 1 GHz (-30%)
Max. current 500 mARMS, 100 A peak
Input Aperture ≥0.065" (1.6 mm)
Voltage Rating 1000 V
Insertion Impedance Terminated: 1 Ω // 5 μH, Un-terminated: 2 Ω // 5 μH
Capacitive Loading 0.5 to 1.5 pF

Pictures