Electrolytic repair 1.jpg | Remove defective cap
Electrolytic repair 1.jpg | Remove defective cap
Electrolytic repair 2.jpg | Cut can near base, remove guts
Electrolytic repair 2.jpg | Cut can near base (using a small hacksaw or a pipe cutter), remove guts
Electrolytic repair 3.jpg | Fit modern cap
Electrolytic repair 3.jpg | Fit modern cap
Electrolytic repair 4.jpg | Epoxy cap onto base
Electrolytic repair 4.jpg | Epoxy cap onto base
Line 45: Line 45:
</gallery>
</gallery>
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:45, 4 May 2019 (PDT)
:--[[User:Peter|Peter]] ([[User talk:Peter|talk]]) 02:45, 4 May 2019 (PDT)
==Wavy trace ==
:'''Symptom:''' Wavy trace
:'''Root cause:''' C786 (100 µF 25 V) has high internal resistance (ESR ~ 20 Ω) causing +12.6 V regulator circuit to oscillate
:'''Solution:''' Replace C786 with new part.
:<gallery>
585 wavy trace.JPG | wavy trace
585_wavy_trace_new_electrolytics.JPG |  +12.6 v PSU with  new eletrolytics.
</gallery>
:--[[User:Matt|Matt]] ([[User talk:Matt|Matt]]) 08:53, 5 May 2019 (PDT)