AA5001: Difference between revisions

532 bytes added ,  10 June 2022
added explanantion about SINAD measurement in DA4084
No edit summary
(added explanantion about SINAD measurement in DA4084)
Line 16: Line 16:
It is intended to be used in conjunction with the [[SG5010]].
It is intended to be used in conjunction with the [[SG5010]].


The '''DA4084''' is a modified variant of the AA5001 that replaces IM measurement with a SINAD function for radio equipment maintenance and characterization.
The '''DA4084''' is a modified variant of the AA5001 that replaces IM measurement with a SINAD function for radio equipment maintenance and characterization. The THD+N measurement of the regular AA5001 fails to measure signals with more than 10% noise content accurately because the autotuning of the notch filter will fail at these high noise levels. The SINAD function in the DA4084 performs the same measurement as THD+N but uses additional circuitry to lock the internal notch filter onto the audio tone at a low noise signal, and then maintains this lock for high noise signals up to 50%, which is sufficient for normal SINAD testing to find radio receivers sensitivity levels.
   
The DA4084 was supplied in a modified [[TM5003]] rack with an unusable left slot and a linear regulating power supply, but also works in generic TM5000 mainframes.  
The DA4084 was supplied in a modified [[TM5003]] rack with an unusable left slot and a linear regulating power supply, but also works in generic TM5000 mainframes.  
The Army designation for this combination  (DA4084 special power supply) is TS-4084/G.
The Army designation for this combination  (DA4084 and special power supply) is TS-4084/G.


{{MissingSpecs}}
{{MissingSpecs}}