P6057: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 1: | Line 1: | ||
{{Probe Sidebar | |||
|manufacturer=Tektronix | |||
|model=P6057 | |||
|summary=1.4 GHz passive ×100 probe | |||
|image=P6056.jpg | |||
|caption=P6056 (otherwise identical ×10 sister model) | |||
|for=50 Ω scope inputs | |||
|introduced=1972 | |||
|discontinued=(?) | |||
|manuals= | |||
* [[Media:010-6056-03.pdf|Tektronix P6056/P6057 Manual]] | |||
}} | |||
The '''Tektronix P6057''' is a 1.4 GHz ×100 probe for 50 Ω-input scopes, released along with the [[7A19]] amplifier in 1972. | The '''Tektronix P6057''' is a 1.4 GHz ×100 probe for 50 Ω-input scopes, released along with the [[7A19]] amplifier in 1972. | ||
Line 15: | Line 26: | ||
{{Spec | Features | Readout Pin, ID Button }} | {{Spec | Features | Readout Pin, ID Button }} | ||
{{EndSpecs}} | {{EndSpecs}} | ||
<blockquote> | <blockquote> | ||
<u>WARNING</u>: | <u>WARNING</u>: | ||
This high performance device has a fragile resistor assembly in the | This high performance device has a fragile resistor assembly in the probe head, subject to failure in rough usage.<br /> | ||
probe head, subject to failure in rough usage. | Use with slip-on tip whenever possible. One replacement part is provided with the package.<br /> | ||
whenever possible. One replacement part is provided with the package.<br /> | Contact nearest Field Office for additional parts (P/N 307-0330-00), if required.<br /> | ||
Contact nearest Field Office for additional parts (P/N 307-0330-00), if<br /> | For replacement and Calibration Procedure, refer to Pg. 3 & 4 of manual.<br /> | ||
</blockquote> | </blockquote> | ||
{{MissingPics}} | |||
[[Category:50 Ω Oscilloscope probes]] | [[Category:50 Ω Oscilloscope probes]] | ||
Latest revision as of 02:04, 21 July 2022
The Tektronix P6057 is a 1.4 GHz ×100 probe for 50 Ω-input scopes, released along with the 7A19 amplifier in 1972.
It has an Identify switch and readout contact.
Its sister model P6056 has ×10 attenuation.
Key Specifications
Bandwidth | DC to 1.4 GHz |
---|---|
Rise time | 250 ps |
Attenuation | 100× |
Input impedance | 5 kΩ // approx. 1 pF |
Scope input impedance | 50 Ω |
Maximum input Voltage | 50 V dc or RMS |
Features | Readout Pin, ID Button |
WARNING:
This high performance device has a fragile resistor assembly in the probe head, subject to failure in rough usage.
Use with slip-on tip whenever possible. One replacement part is provided with the package.
Contact nearest Field Office for additional parts (P/N 307-0330-00), if required.
For replacement and Calibration Procedure, refer to Pg. 3 & 4 of manual.
Pictures
- please add