It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed | It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CCD output signals at 4 MS/s. | ||
<blockquote> | <blockquote> |
Revision as of 12:33, 12 March 2019
The Tektronix 2440 is a two-channel 300 MHz, 500 MS/s digital sampling oscilloscope from the late 1980s.
It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CCD output signals at 4 MS/s.
WARNING:
Operation of the 2440 (and 2432 and 2430): A/D-converter and CCD hybrids may suffer irreparable damage within a few minutes if operated out of the metal enclosure without extra cooling.
Key Specifications
Bandwidth | 300 MHz (Limit: 20, 100, Full) |
---|---|
Sampling rate | 500 MSa/s each channel with simultaneous acquisition |
Deflection | 2 mV/div to 5 V/div in 1-2-5 sequence |
Auto Scale Factor | On-Screen Readout indicates probe tip deflection factors for ×1, ×10, ×100 and ×1000 probes |
Input Impedance | 1 MΩ // 15 pF |
Sweep Speeds | 2 ns/div to 5 s/div in 1-2-5 sequence (Modes A, A Intensified and B) |
Auto Measure | Simultaneously measures up to four parameters on one or more displayed waveforms, updated at 3 Hz intervals or display a snapshot of 20 parameters of waveform target |
Acquisition Modes | Normal, Envelope (glitch capture), Average |
Display Modes | CH1, CH2, invert, add, multiply, X-Y |
Cursors | Volts, Time, Volts At Time, 1/Time, Slope |
Functions | A Mode - Auto Level, Auto/Roll, Normal and Single Seq; B Mode - Trigger After Delay, Runs After Delay, 1K Record Length and more |
Interface | GPIB |
Probes | Two passive P6137 400 MHz ×10 voltage probes, 1.3 m |
Power | 90 to 132 VAC or 80 to 250 VAC, 48 to 440 Hz; 160 W |
Features |
|
Options |
|
Links
Internals
The 2440 employs several processors − an MC68B09 main CPU controls the instrument, a custom 16-bit waveform processor performs signal computations, and an MC6805 microcontroller handles the front panel interface. The GPIB Interface is implemented with a TMS994 controller.
The power supply consists of a buck-type pre-regulator followed by a pus-pull inverter and linear post-regulators (except for the +5 V digital supply rail).
CRT is a 154-0850-01.