The '''Tektronix 2440''' is a two-channel 300 MHz, 500 MS/s digital sampling oscilloscope from the late 1980s.
The '''Tektronix 2440''' is a two-channel 300 MHz, 500 MS/s digital sampling oscilloscope from the late 1980s.


It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CD output signals at 4 MS/s.   
It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CCD output signals at 4 MS/s.   


<blockquote>
<blockquote>

Revision as of 12:33, 12 March 2019

{{{manufacturer}}} 
300 MHz, 500 MS/s 2-ch digital scope
Tektronix 2440

Produced from 1988 to (?)

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix 2440 is a two-channel 300 MHz, 500 MS/s digital sampling oscilloscope from the late 1980s.

It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CCD output signals at 4 MS/s.

WARNING:
Operation of the 2440 (and 2432 and 2430): A/D-converter and CCD hybrids may suffer irreparable damage within a few minutes if operated out of the metal enclosure without extra cooling.

Key Specifications

Bandwidth 300 MHz (Limit: 20, 100, Full)
Sampling rate 500 MSa/s each channel with simultaneous acquisition
Deflection 2 mV/div to 5 V/div in 1-2-5 sequence
Auto Scale Factor On-Screen Readout indicates probe tip deflection factors for ×1, ×10, ×100 and ×1000 probes
Input Impedance 1 MΩ // 15 pF
Sweep Speeds 2 ns/div to 5 s/div in 1-2-5 sequence (Modes A, A Intensified and B)
Auto Measure Simultaneously measures up to four parameters on one or more displayed waveforms, updated at 3 Hz intervals or display a snapshot of 20 parameters of waveform target
Acquisition Modes Normal, Envelope (glitch capture), Average
Display Modes CH1, CH2, invert, add, multiply, X-Y
Cursors Volts, Time, Volts At Time, 1/Time, Slope
Functions A Mode - Auto Level, Auto/Roll, Normal and Single Seq; B Mode - Trigger After Delay, Runs After Delay, 1K Record Length and more
Interface GPIB
Probes Two passive P6137 400 MHz ×10 voltage probes, 1.3 m
Power 90 to 132 VAC or 80 to 250 VAC, 48 to 440 Hz; 160 W
Features
  • Autoprobe Function With P6137 Probe
  • 300 MHz X-Y bandwidth at probe tip
  • Automatic measurements
  • extensive triggering capabilities
  • Printer/Plotter Output
  • On-screen HELP text describing every front-panel control
Options

Links

Internals

The 2440 employs several processors − an MC68B09 main CPU controls the instrument, a custom 16-bit waveform processor performs signal computations, and an MC6805 microcontroller handles the front panel interface. The GPIB Interface is implemented with a TMS994 controller.

The power supply consists of a buck-type pre-regulator followed by a pus-pull inverter and linear post-regulators (except for the +5 V digital supply rail).

CRT is a 154-0850-01.

Pictures