5T1A: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 10: | Line 10: | ||
|designers= | |designers= | ||
|manuals= | |manuals= | ||
* [[Media:070-387.pdf|Tektronix 5T1A Manual | * [[Media:070-387.pdf|Tektronix 5T1A Manual]] | ||
* [ | * [https://w140.com/tek_fcp/tek_type_5t1a_factory_cal_proc.pdf Tektronix 5T1A Field Recalibration Procedure] | ||
* [[Media:050-0419-00.pdf|Modification Note Documenting Replacement of 152-0074-00 Tunnel Diode with 152-0381-00 | * [[Media:050-0419-00.pdf|Modification Note Documenting Replacement of 152-0074-00 Tunnel Diode with 152-0381-00]] | ||
* [[Media:Tek 5t1a irb no ocr.pdf|Tektronix 5T1A Instrument Reference Book ( | * [[Media:Tek 5t1a irb no ocr.pdf|Tektronix 5T1A Instrument Reference Book]] (needs OCR) | ||
* [[Media:Tek 5t1a cal outline.pdf|Tektronix 5T1A Calibration Outline ( | * [[Media:Tek 5t1a cal outline.pdf|Tektronix 5T1A Calibration Outline]] (OCR) | ||
* [[Media:Tek 5t1a preliminary.pdf|Tektronix 5T1A Preliminary Manual]] | * [[Media:Tek 5t1a preliminary.pdf|Tektronix 5T1A Preliminary Manual]] | ||
}} | }} | ||
The '''Tektronix 5T1A''' is a timing plug-in [[introduced in 1963]] for the [[661]] [[sampling oscilloscope]]. | The '''Tektronix 5T1A''' is a timing plug-in [[introduced in 1963]] for the [[661]] [[sampling oscilloscope]]. It is an improved version of the [[5T1]]. | ||
The job of the 5T1A is to produce three signals – the sampling pulse, the horizontal sweep, and the blanking signal. | |||
The input to the 5T1A is a trigger signal. The trigger can arrive from any of four sources: | The input to the 5T1A is a trigger signal. The trigger can arrive from any of four sources: | ||
* internally from the trigger pick-off in a [[4S1]] or [[4S2A]] plug-in | * internally from the trigger pick-off in a [[4S1]] or [[4S2A]] plug-in |
Revision as of 14:30, 4 June 2023
The Tektronix 5T1A is a timing plug-in introduced in 1963 for the 661 sampling oscilloscope. It is an improved version of the 5T1.
The job of the 5T1A is to produce three signals – the sampling pulse, the horizontal sweep, and the blanking signal. The input to the 5T1A is a trigger signal. The trigger can arrive from any of four sources:
- internally from the trigger pick-off in a 4S1 or 4S2A plug-in
- externally via a GR-874 connector on the front panel of the 5T1A
- from the calibration oscillator in the 661, via the multi-pin plug-in connector
- free-running
Each of these modes is suited to some measurement scenarios. In the case of internal triggering, the signal goes from the trigger pick-off in the 4S1 to a coaxial interconnect that passes through the 661, connecting the sampling unit to the timing unit.
Triggering in the 5T1A is accomplished using five 1N3129 20 mA tunnel diodes.