7000-series remote control connector: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 1: Line 1:
The '''remote control connector''' on the rear of the [[7904]] (connector J90) and [[7503]]/[[7503]] scopes (connector J1075) is an [[Amphenol 165 series connectors|Amphenol 165-16]] socket.
The '''remote control connector''' on the rear of the [[7904]] (connector J90), [[7503]]/[[7504]], and [[7514]] scopes (connector J1075) is an [[Amphenol 165 series connectors|Amphenol 165-16]] socket.


{| class="wikitable"
{| class="wikitable"
Line 19: Line 19:
| F || Remote single-shot lock-out (**) || Closure to ground within 0.4 V, pin E was grounded before, rate of change ≥ 0.1 V/μs, 3 mA max, 10 V open voltage
| F || Remote single-shot lock-out (**) || Closure to ground within 0.4 V, pin E was grounded before, rate of change ≥ 0.1 V/μs, 3 mA max, 10 V open voltage
|-
|-
| H || n/c ||
| H || Upper screen remote erase (***) || Closure to ground within 0.4 V, rate of change at least 1 V/μs
|-
|-
| J || n/c ||
| J || Lower screen remote erase (***) || Closure to ground within 0.4 V, rate of change at least 1 V/μs
|-
|-
| K || n/c ||
| K || Remote write thru (***) || Closure to ground within 0.4 V, rate of change at least 1 V/μs
|-
|-
|}
|}
Line 29: Line 29:
:'' (*) not in 7503''
:'' (*) not in 7503''
:'' (**) only in 7904''
:'' (**) only in 7904''
:'' (***) only in [[7514]]''


[[Category:Instrument specific interfaces]]
[[Category:Instrument specific interfaces]]

Revision as of 13:18, 10 November 2021

The remote control connector on the rear of the 7904 (connector J90), 7503/7504, and 7514 scopes (connector J1075) is an Amphenol 165-16 socket.

Pin Function Notes
A Single sweep reset Close to ground within 0.5 V, 10 mA max, minimum 10 μs pulse width
B Ground
C A Remote ready +5 V via 47 Ω when active, can drive #49 bulb
D B Remote ready (*) +5 V via 47 Ω when active, can drive #49 bulb
E Remote readout lock-out (**) Close to ground within 0.1 V, 2 mA max, 2 V open voltage
F Remote single-shot lock-out (**) Closure to ground within 0.4 V, pin E was grounded before, rate of change ≥ 0.1 V/μs, 3 mA max, 10 V open voltage
H Upper screen remote erase (***) Closure to ground within 0.4 V, rate of change at least 1 V/μs
J Lower screen remote erase (***) Closure to ground within 0.4 V, rate of change at least 1 V/μs
K Remote write thru (***) Closure to ground within 0.4 V, rate of change at least 1 V/μs
(*) not in 7503
(**) only in 7904
(***) only in 7514