P6055: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 31: | Line 31: | ||
{{EndSpecs}} | {{EndSpecs}} | ||
==Links== | |||
{{Documents|Link=P6055}} | |||
{{MissingPics}} | |||
[[Category:1 MΩ Oscilloscope probes]] | [[Category:1 MΩ Oscilloscope probes]] | ||
[[Category:Passive Differential probes]] | [[Category:Passive Differential probes]] |
Latest revision as of 06:06, 12 October 2023
The Tektronix P6055 is a 10× probe pair for differential amplifiers such as e.g. the 7A13 and 5A22N having 1 MΩ Inputs and 20 to 47 pF input capacity.
The probe's resistive attenuation ratio is adjustable to an exact ×10 for the given amplifier input. This allows precise gain matching to improve CMRR.
Tektronix also recommended the P6055 for the 7D12/M2 sample/hold, because of the trimming feature improving the accuracy of voltage readings.
The P6055 has a readout pin.
Key Specifications
Attenuation | Adjustable to ×10 |
---|---|
Input impedance | 1 MΩ // 10-12.5 pF |
Step response | < 4 ns |
Compensation range | 1 MΩ//12-47 pF |
CMRR | with 7A13 or 1A5, 20,000:1 from DC – 1kHz, 100:1 at 20 MHz |
Max. input | ±500 V maximum common-mode |
Links
Documents Referencing P6055
Document | Class | Title | Authors | Year | Links |
---|---|---|---|---|---|
Tekscope 1972 V4 N5 Sep 1972.pdf | Article | A Practical Approach to Differential Amplifiers and Measurements | Fred Beckett | 1972 | 112 • 11A33 • 1A5 • 1A6 • 1A7 • 1A7A • 26A2 • 3A9 • 5A22N • 7A13 • 7A22 • AM502 • D • E • G • P6055 • P6046 • W • Z |
Tekscope 1972 V4 N6 Nov 1972.pdf | Article | Differential Amplifiers and Measurements, Part 2 | Fred Beckett | 1972 | 112 • 11A33 • 1A5 • 1A6 • 1A7 • 1A7A • 26A2 • 3A9 • 5A22N • 7A13 • 7A22 • AM502 • D • E • G • P6055 • P6046 • W • Z |
Tekscope 1975 V7 N5.pdf | Article | Replacing Probe Tips, Bodies, and Boots | Ron Lang | 1975 | P6049A • P6049B • P6053A • P6053B • P6054A • P6054B • P6055 • P6062A • P6063A • P6065A • P6075A |
Pictures
- please add