P6135: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 2: | Line 2: | ||
|manufacturer=Tektronix | |manufacturer=Tektronix | ||
|model=P6135 | |model=P6135 | ||
|summary= | |summary=150 MHz, ×10 differential probe pair | ||
|image=Tek p6135 1.jpg | |image=Tek p6135 1.jpg | ||
|caption=Tektronix P6135 | |caption=Tektronix P6135 |
Revision as of 04:20, 18 July 2024
The Tektronix P6135(A) is a differential probe pair matched to minimize differences in impedance and signal delay, to maintain CMRR up to 20,000:1 when used with differential amplifiers such as the 7A13, 7A22 or 11A33.
The P6135 is a subminiature modular probe with a limited compensation range around 15 pF. The P6135A is a larger compact modular probe and comes with three different tips for increased compensation range, matching the 11A33 − red, 7A13 − black, and 7A22 − blue. The P6135A model can therefore be seen as a replacement of the P6055.
Key Specifications
Scope Impedance Compensation Range | P6135: 1 MΩ // 13-17 pF; P6135A: 1 MΩ // 13-17 pF (red tip), 1 MΩ // 18-22 pF (black tip) and 1 MΩ // 45-49 pF (blue tip) |
---|---|
Attenuation | Adjustable to ×10 |
CMRR | 10,000:1 on 11A33, 20,000:1 on 7A13 |
Bandwidth | 150 MHz |
Max. input | ±500 V Maximum Common Mode Voltage |