SA501: Difference between revisions
No edit summary |
No edit summary |
||
Line 7: | Line 7: | ||
It requires a [[P6405]] probe for the data input, and a [[P6458]] 4-channel probe for control inputs (clock, start, stop, qualifier). The P6458 can be substituted by the similar, more common [[P6451|P6451 8+1 channel probe]] without increase in functionality. | It requires a [[P6405]] probe for the data input, and a [[P6458]] 4-channel probe for control inputs (clock, start, stop, qualifier). The P6458 can be substituted by the similar, more common [[P6451|P6451 8+1 channel probe]] without increase in functionality. | ||
{{ | {{BeginSpecs}} | ||
{{Spec|Data Input | 1 MΩ to ground or 50 kΩ to 1.4 V, switchable; 14 pF; max. 40 V DC+peak AC}} | |||
{{Spec|Threshold | fixed TTL internal, or variable ±12 V }} | |||
{{Spec|Data rate | 40 MHz max.; 15 ns setup time, 0 ns hold time }} | |||
{{Spec|CRC polynomial | x<sup>16</sup>+x<sup>12</sup>+x<sup>9</sup>+x<sup>7</sup>+1 }} | |||
{{Spec|Display | 4-digit modified hexadecimal (0-9,A,C,F,H,P,U) }} | |||
{{EndSpecs}} | |||
==Links== | ==Links== |
Revision as of 04:46, 13 October 2024
The Tektronix SA501 is a signature analyzer plug-in for the TM500 system.
It is also known as the 067-1090-00.
It requires a P6405 probe for the data input, and a P6458 4-channel probe for control inputs (clock, start, stop, qualifier). The P6458 can be substituted by the similar, more common P6451 8+1 channel probe without increase in functionality.
Key Specifications
Data Input | 1 MΩ to ground or 50 kΩ to 1.4 V, switchable; 14 pF; max. 40 V DC+peak AC |
---|---|
Threshold | fixed TTL internal, or variable ±12 V |
Data rate | 40 MHz max.; 15 ns setup time, 0 ns hold time |
CRC polynomial | x16+x12+x9+x7+1 |
Display | 4-digit modified hexadecimal (0-9,A,C,F,H,P,U) |
Links
Documents Referencing SA501
- (no results)
Patents that may apply to SA501
Page | Title | Inventors | Filing date | Grant date | Links |
---|---|---|---|---|---|
Patent US 3976864A | Apparatus and method for testing digital circuits | Gary B. Gordon • George A. Haag • Jan R. Hofland • Daniel I. Kolody | 1974-09-03 | 1976-08-24 | Signature analysis • 308 • SA501 |
Internals
The SA501 is implemented in standard ECL and TTL logic without a microprocessor. A linear feed back shift register accumulates the CRC. The display is multiplexed, with an IM5610 32×8 PROM used as a 7-segment decoder.
It contains a switch-mode converter fed off the +33.5 V supply to create the internal +5 V rail.
The manual includes signatures for troubleshooting a SA501 with another signature analyzer.
Pictures
067-1090-00
-
-
-
-
Note that the P6407 probe shown connected to the SA501 here is not compatible and can be damaged when the SA501 is powered.
-
-
-
-
-
-
-
SA 501
Components
Some Parts Used in the SA501
Part | Part Number(s) | Class | Description | Used in |
---|---|---|---|---|
IM5610 | Monolithic integrated circuit | 32×8 (256 bit) PROM | 7912AD • MicroLab I • SA501 |