The Tektronix 492 is a spectrum analyzer with a frequency range of 10 kHz to 21 GHz in coax, and up to 325 GHz with external waveguide mixers (492PGM N/A). The P suffix designation indicates GPIB Programmability.

During the lifespan of the instrument, the specifications and included options were altered several times. A major upgrade was the introduction of the 492A in 1987 which added marker functionality and resolution bandwidths and the 492BP in 1989 which added a counter and increased the displayed dynamic range. The 492PGM is a cost reduced version introduced in 1990. Also see 49X Series Comparison.

Key Specifications

Frequency
  • Coaxial Input Bands: 10 kHz to 21 GHz
  • External Waveguide mixers: 10 kHz to 325 GHz (applies to 492BP not 492PGM)
Frequency Span 492BP: 100 Hz/div to 10 GHz/div; 492PGM: 200 MHz to 1 GHz; both in 1, 2, 5 sequence; plus 0 Hz and MAX
Resolution Bandwidth (-6 dB bandwidth) 492BP: 100 Hz to 3 MHz; 492PGM: 1 kHz to 3 MHz; both in decade steps
RF Input 50 Ω, max. +30 dBm (1 W) CW; / 75 W peak pulse (1 µs, 0.1% duty factor)
RF Attenuator 0 dB to 60 dB, 10 dB steps
Reference Level -117 dBm to +30 dBm
Sweep Speed 10 sec/div to 20 µs/div, 1−2−5
Video Bandwidth 492BP: 0.3 Hz to 30 kHz; 492PGM: 3 Hz to 30 kHz
Memory NVRAM for up to 9 waveform displays, up to 10 front panel setups, plus 8 K for programming macros
Triggering Modes Free Run, Line, Video, Single, External
Displayed Average Noise −30 dBm to −131 dBm
Display Dynamic Range 492BP: 90 dB; 492PGM: 80 dB
Calibrator (Cal out) 50 Ω, -20 dBm ±0.3 dB at 100 MHz
Weight 492BP: 21.8 kg (47 lbs) 492PGM: 21.3 kg (46 lbs)
Power 90 − 132 VAC, 48 to 440 Hz; 180 – 250 VAC, 48 to 440 Hz. At 115 VAC, 60 Hz, 210 W max

Options

  • Opt. 01: Internal Preselection. (Limits the first band to 1.8 GHz instead 4.2 GHz). Later became standard
  • Opt. 02: Digital Storage. Later became standard
  • Opt. 03: Frequency Stabilization & 100 Hz Resolution. Later became standard
  • Opt. 07: 75 Ω dBmV input and calibration in addition to the normal 50 Ω dBm input and calibration. (Not combinable with Opt. 21 and 22; no external mixer capability). Include 42-inch 75 Ω BNC-BNC coax cable and BNC male to “F” female adapter.
  • Opt. 21: (492BP) High performance 18 to 40 GHz WM490 Series Waveguide Mixer Set
  • Opt. 22: (492BP) Same as Opt. 21 plus WM490U (40-60 GHz) Waveguide Mixer
  • Opt. 23: GRASP software, PC2A interface, and GPIB cable
  • Opt. 27: Epson LT-386SX, GRASP software, PC2A interface, and GPIB cable
  • Opt. 28: Compaq Deskpro 386S, Model 40, GRASP software, PC2A interface, and GPIB cable
  • Opt. 39: Non-lithium (Silver) batteries for battery-backed memory
  • Opt. 41: Digital Microwave Radio Measurement Enhancement package
  • Opt. 42: Replaces MARKER/VIDEO input port on the rear panel with a 110 MHz IF output port that provides a 3 dB signal bandwidth ≥ 4.5 MHz
  • Opt. 45: (all except 492PGM) MATE/CIIL language interface

Links

Documents Referencing 308

Document Class Title Authors Year Links
Tekscope 1980 V12 N1.pdf Article New Products 1980
Tekscope 1980 V12 N2.pdf Article Portable Analyzer Speeds Test and Service of Microprocessor-Based System Ed Averill John Huber 1980

Connections

Front panel inputs/outputs are:

  • RF INPUT (50 Ω N connector) – Input for RF signals to 21 GHz. If input signal has a DC component, use a blocking capacitor
  • CAL OUT (BNC connector) – Provides 100 MHz, –20 dBm signal and a comb of frequency markers 100 MHz apart
  • 1ST LO OUTPUT (SMA connector) – Output of the 1st local oscillator, must be terminated into 50 Ω when not connected.
  • 2ND LO OUTPUT (SMA connector) – Output of the 2nd local oscillator, must be terminated into 50 Ω when not connected.
  • EXTERNAL MIXER RF INPUT (TNC connector, 50 Ω)
  • Camera Power

Rear panel inputs/outputs are:

  • PROBE POWER – Provides operating voltages (+5 V, -15 V, +15 V; 100 mA max each) for active probes
  • HORIZ/TRIG (input, BNC) – In External Triggering mode, AC coupled input for trigger signals; when TIME/DIV selection is EXT, DC coupled input for horizontal sweep voltages
  • MARKER/VIDEO (B053575+, output, BNC) – interfaces the 492/492P with a TV Sideband Adapter, such as the Tek 1405, so a marker from the adapter is displayed on the internal video. External video applied to this connector will also be displayed if pin 1 of the ACCESSORIES connector is grounded.
    • Opt. 42 replaces MARKER/VIDEO with 110 MHz IF (output, BNC) – 3 dB signal bandwidth ≥ 4.5 MHz
  • EXT PRESELECTOR (B053574 and below) (output, BNC) – variable voltage (approx -0.2 V to +10 V) proportional to center frequency, for preselector bands of Opt. 01 instruments, to drive an external preselector
  • HORIZ (output, BNC) – 0.5 V/div horizontal signal
  • VERT (output, BNC) – Video signal with 0.5 V/Div
  • PEN LIFT (output, BNC) – TTL compatible signal to lift the pen of a chart recorder during sweep retrace. In Opt. 42 instruments this port may also be used for inputting external video if pin 1 of the ACCESSORY connector is grounded.
  • 10 MHz IF (output, BNC) – Output of the 10 MHz IF
  • J104 ACCESSORY (B053575+ only), female DB25 connector – Provides bidirectional access to the instrument bus.
  • IEEE STD 488 PORT GPIB interface – Added to all "P" suffixed instruments

Pictures

492

492A

492BP