P6046: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(image)
No edit summary
 
(25 intermediate revisions by 4 users not shown)
Line 1: Line 1:
The Tektronix P6046 is a differential probe and amplifier.
{{Probe Sidebar
It is mainly intended for use with a [[1A5]].
|manufacturer=Tektronix
With a 1A5, it has a 45MHz bandwidth.
|model=P6046
The raw bandwidth of the probe and amplifier is 100MHz.
|summary=100 MHz differential probe  
|image=Tek P6046 catalog 2.jpg
|caption=Tektronix P6046 Differential Probe and 015-0106-00 Amplifier
|for=50 Ω scope inputs
|introduced=1968
|discontinued=1992
|designers=Glenn Bateman
|manuals=
There are three versions of the manual:
* [[Media:070-0756-00.pdf|070-0756-00 - 1968]] − assumes [[1A5]] for absolute gain calibration
* [[Media:070-7129-00.pdf|070-7129-00 - 1988]]
* [[Media:070-7129-01.pdf|070-7129-01 - 1992]]


* [http://w140.com/Tektronix_P6046_Instruction_Manual.pdf P6046 Manual (no schematics, PDF)]
* [[Media:P6046 Voltage and Waveform Information.pdf|P6046 Voltage and Waveform Information]]
* [http://w140.com/Tek_P6046_Differential_Probe-Operation_Manual.pdf P6046 Detailed Manual (no schematics, PDF)]
* [[Media:P6046 Schematics Rev. C June 1977 11x17.pdf|P6046 Schematics Rev. C June 1977]]
* [http://w140.com/P6046_8.5x11.pdf P6046 Voltage and Waveform Information (PDF)]
* [http://w140.com/P6046_17x11.pdf P6046 Schematics (PDF)]


* [[Media:Tek p6046 eis signature page.pdf|P6046 Engineering Instrument Specification Signature Page]]
* [[Media:Tek p6046 fcp march 1968.pdf|P6046 Factory Calibration Procedure, March 1968]]
}}
The '''Tektronix P6046''' is a 100 MHz differential probe oscilloscope probe [[introduced in 1968]] and discontinued sometime after 1992.
It was designed by [[Glenn Bateman]].
It was initially intended for use with a [[1A5|1A5 plug-in]], to which it attaches through a 9-pin [[Amphenol 165 series connector]].
A separate amplifier (015-0106-00) was available as an option to use the probe on a 50 Ω input scope set at 10 mV/Div.
An additional ×10 attenuator (010-0361-00) can be plugged onto the P6046's inputs.
Note that the probe itself acts as either a 1× or 10× attenuation probe dependent on a signal voltage on pin D of the connector. 
When used with the amplifier, 10× mode is used in the 200, 100, and 50 mV/div settings and 1× mode for 20, 10, 5, 2, and 1 mV/div.
{{BeginSpecs}}
{{Spec | Bandwidth | 45 MHz with 1A5; 100 MHz with amplifier; AC low frequency cut-off 20 Hz (2 Hz with ×10 head) }}
{{Spec | Input impedance | 1 MΩ // 10 pF (with ×10 head, 10 MΩ // 3 pF) }}
{{Spec | Common-mode range | ±5 V DC+peak AC up to 10 MHz, decreasing to ±2 V at 50 MHz }}
{{Spec | Maximum Allowable Input | 25 V total DC+peak AC, 25 V total difference between + and − input tips }}
{{EndSpecs}}
==Operation==
The two probe tips are (in DC mode) directly connected to the gates of a pair of JFETs. 
Use the shorting plugs on the tips when not in use:
<blockquote>Ground the probe!  Seriously.  Sticker the probe "Ground Me!"  I also recommend stickering the probe amplifier with the ATTEN BAL setup procedure from section 2-6 of the manual.</blockquote>
==Links==
* [http://www.i9t.net/p6046/p6046.html P6046 @ iceNINE Tech]
* [https://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_other/P6046.htm P6046 @ amplifier.cd]
{{Documents|Link=P6046}}
==Pictures==
<gallery>
<gallery>
File:P6046.jpg
P6046_1.jpg | P6046
File:Tek p6046 internal1.jpg
P6046_2.jpg | P6046
File:Tek p6046 internal2.jpg
Tek p6046 internal1.jpg
File:Tek p6046 coupling.jpg
Tek p6046 internal2.jpg
Tek p6046 coupling.jpg
015-0106-00.jpg | 015-0106-00 amplifier and power supply
</gallery>
</gallery>
[[Category:Differential oscilloscope probes]]

Latest revision as of 06:14, 28 January 2024

Tektronix P6046
100 MHz differential probe
Tektronix P6046 Differential Probe and 015-0106-00 Amplifier

Compatible with 50 Ω scope inputs

Produced from 1968 to 1992

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6046 is a 100 MHz differential probe oscilloscope probe introduced in 1968 and discontinued sometime after 1992. It was designed by Glenn Bateman.

It was initially intended for use with a 1A5 plug-in, to which it attaches through a 9-pin Amphenol 165 series connector.

A separate amplifier (015-0106-00) was available as an option to use the probe on a 50 Ω input scope set at 10 mV/Div.

An additional ×10 attenuator (010-0361-00) can be plugged onto the P6046's inputs.

Note that the probe itself acts as either a 1× or 10× attenuation probe dependent on a signal voltage on pin D of the connector. When used with the amplifier, 10× mode is used in the 200, 100, and 50 mV/div settings and 1× mode for 20, 10, 5, 2, and 1 mV/div.

Key Specifications

Bandwidth 45 MHz with 1A5; 100 MHz with amplifier; AC low frequency cut-off 20 Hz (2 Hz with ×10 head)
Input impedance 1 MΩ // 10 pF (with ×10 head, 10 MΩ // 3 pF)
Common-mode range ±5 V DC+peak AC up to 10 MHz, decreasing to ±2 V at 50 MHz
Maximum Allowable Input 25 V total DC+peak AC, 25 V total difference between + and − input tips

Operation

The two probe tips are (in DC mode) directly connected to the gates of a pair of JFETs.

Use the shorting plugs on the tips when not in use:

Ground the probe! Seriously. Sticker the probe "Ground Me!" I also recommend stickering the probe amplifier with the ATTEN BAL setup procedure from section 2-6 of the manual.

Links

Documents Referencing P6046

Document Class Title Authors Year Links
Service Scope 46 Oct 1967.pdf Article P6046 DC-to-100 MHz Differential Probe and Amplifier Glenn Bateman 1967
Tekscope 1972 V4 N5 Sep 1972.pdf Article A Practical Approach to Differential Amplifiers and Measurements Fred Beckett 1972
Tekscope 1972 V4 N6 Nov 1972.pdf Article Differential Amplifiers and Measurements, Part 2 Fred Beckett 1972

Pictures