Tektronix 3S76
Dual Channel Sampling plug-in
Tektronix 3S76 front

Compatible with 560-series scopes

Produced from 1962 to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 3S76 is a dual-trace sampling plug-in for 560-series scopes. The inputs have 50 Ω impedance and use GR-874 connectors. Rise time is rated at less than 0.4 ns.

The initial design work on the 3S76 was done by Cliff Moulton. It was called the Type 76, later renamed 3S76 before going into production. A decision was made to demonstrate the 76/3S76 at the 1961 IRE show. This created a panic because the plug-in was not complete. The 76 project landed in the TV group. Ron Olson and Charlie Rhodes were designing it and Walt Lowy was a technician. Phil Crosby worked on it for a week during spring break from Portland State (college then, not university). Phil Crosby designed the channel switching circuitry.

Key Specifications

Rise time < 400 ps
Bandwidth 875 MHz
Input impedance 50 Ω
Triggering Internal trigger take-off, 55 ns delay line

Early 3S76 units are marked RT. ≈ 0.4 nSEC on the front panel. Later 3S76 units are marked RT. ≤ 0.4 nSEC on the front panel.

Internals

Each input has a trigger takeoff transformer and a 55 ns coaxial delay line. The trigger source channel is selectable.

Each delay line terminates at that channel's sampling bridge, which is a four-diode design using beam-lead GaAs diodes.

A 7586 nuvistor buffers the output of each sampling bridge. The sampling pulse, common to both channels, is generated by a 152-115 snap-off diode and a 3 cm twisted-pair clipping line.