067-0657-00
The Tektronix 067-0657-00 Normalized Ramp Generator is a calibration plug-in for 7000-series scopes.
It provides calibrated sweep rates from 0.5 ns/Div to 50 ns/Div (in a 1–2–5 sequence) plus 1 μs/Div and 1 ms/Div.
The calibration fixture is triggered from an external signal (such as a trigger output from a time-mark generator) applied to the front-panel TRIG IN connector.
Internals
The 067-0657-00 was the result of some fundamental research by Bruce Hofer inspired by production issues in the calibration of early 7904 scopes. This lead to a new sweep circuit that uses current steering in a differential pair to start the charging of the timing capacitor as is discussed in patent 4009399. The sweep circuit in the 067-0657-00 follows the patent application but does not implement the baseline stability circuit, which was incorporated in the 7B92A.
Links
Pictures
Components
Some Parts Used in the 067-0657-00
Part | Part Number(s) | Class | Description | Used in |
---|---|---|---|---|
155-0049-00 | 155-0049-00 • 155-0049-01 • 155-0049-02 | Monolithic integrated circuit | sweep control with lockout | 335 • 464 • 465 • 466 • 475 • 475A • 475M • 485 • 5B31 • 5B40 • 5B52 • 5B42 • 5B44 • 7B53A • 7B80 • 7B85 • 7B87 • 7B92A • 7B90P • 7B10 • 7B15 • SC502 • 7B42N • AN/USM-281C • 067-0657-00 |