1401: Difference between revisions

From TekWiki
Jump to navigation Jump to search
m (in progress)
No edit summary
 
(29 intermediate revisions by 6 users not shown)
Line 1: Line 1:
[[Image:Tek1401a-front.JPG|thumb|600px|right|Tektronix 1401A Spectrum Analyzer Module]]
{{Instrument Sidebar
|manufacturer=Tektronix
|model=1401
|class=Spectrum Analyzer
|series=
|summary=Spectrum analyzer front-end
|image=Tek1401a-front.jpg
|caption=Tektronix 1401A Spectrum Analyzer Module
|introduced=1971
|discontinued=(?)
|designers=
|manuals=
* [[Media:070-1071-00.pdf|Tektronix 1401 Manual]]
* [[Media:070-1197-00.pdf|Tektronix 1401A Manual]]
* [[Media:070-1197-00 (1982).pdf|Tektronix 1401A Manual (1982)]]
* [[Media:062-1327-00.pdf|Tektronix 323-1401 Cabinet Assembly]]
}}
The '''Tektronix 1401''' is a spectrum analyzer front-end covering the frequency range of 1 MHz to 500 MHz,
designed to be used with an external oscilloscope, typically a [[323]], [[324]] or [[326]]


It covers the frequency range from 1MHz to 500MHz.
The 1401's case, like the [[1501|1501 TDR]]'s, is the same size as the 323/324 case and designed to be stacked with the scope.
It uses an external oscilloscope,
It has internal rechargeable NiCd batteries and can also operate from AC power (using an unusual power cord) or external DC from 6 V to 16 V.
typically a [[323]] or [[324]],
as a display.
The 1401A can be operated on batteries or 12V or wall power.


See TekScope January 1972 pages 14-15 for some servicing notes that
The external oscilloscope is assumed to have 10 horizontal divisions and 6 vertical divisions.
Tek wrote after the manual was published.
Tektronix says it is compatible with any oscilloscope having 0.5 V/div horizontal deflection factor and 1.2 V full screen vertical deflection. 
Note that most full sized Tektronix oscilloscopes have 8 vertical divisions so use care when doing the initial setup and reading the scale in dBm. 
Also note the lack of a Z-axis output.
 
There appear to be two variants of the 1401A.
Both the manual and catalog specify the resolution bandwidth selector as 1000, 100 and 3 kHz,
but the later versions are 1000, 100, and 10 kHz. 
There is also a 1401A-1 which has a 75 Ω input impedance.
The predecessor of the 1401A, the 1401, did not have the BNC "cal output" or "cal on" pushbutton switch.
 
{{MissingSpecs}}
 
==Links==
* [https://www.youtube.com/watch?v=4Y6ji0QBsww Spectrum Analyzer Basics / Tutorial, and the Tektronix 1401A] by Alan Wolke W2AEW @ YouTube
{{Documents|Link=1401}}
{{PatentLinks|1401}}
 
==Internals==
Operation is typical of a swept front end analyzer.  The input is attenuated and low-pass filtered to 500 MHz. 
It is then mixed with the first LO, a varactor tuned oscillator that can sweep 695 MHz to 1195 MHz. 
The first mixer is a double balanced, four diode star.  The signal is DC coupled to this mixer. 
The manual is not clear as to the damage level; with no attenuation they say that it is non linear above -30 dBm. 
The first IF output is 695 MHz, filtered to remove the other image.
The first IF is mixed with a 720 MHz fixed 2nd LO to produce the 2nd IF at 25 MHz.
 
The 25 MHz 2<sup>nd</sup> IF leaves the shielded RF housing and enters the IF board via SMB connector P210.
This 2<sup>nd</sup> IF is then amplified and mixed with a 30 MHz fixed 3rd LO to produce the 3<sup>rd</sup> IF of 5 MHz. 
This 3<sup>rd</sup> IF is then passed through the selectable RBW filters, detected and send to the video out.
 
==Repair issues==
See [[Media:Tek1401a-tekscopes.pdf|TekScope January 1972 pages 14-15]] for some servicing notes that Tek wrote after the manual was published.


There is one tunnel diode found in the external trigger circuit.
There is one tunnel diode found in the external trigger circuit.
It is a 152-0402-00, a 2mA, 25pF tunnel diode in a DO-17 package.
It is a [[152-0402-00]], a 2 mA, 25 pF tunnel diode in a DO-17 package. It is only used when the external trigger input is connected.
It's only necessary if you use the external trigger input.
 
There is one Tektronix custom IC, [[155-0042-02]], listed as a Miller Integrator.
It is used to generate the sweep ramp voltage.  The package is a 10 pin metal can.
 
The input attenuator is a fairly custom part due to the IF gain pot that is concentric with the RF atten control.
Dissassembly of the attenuator is easy after removing it from the 1401A. 
Note that the IF gain pot is removed via two allen setscrews. 
To bench test the attenuator with DC, use a 50 Ω terminator on SMA input and connect the output SMA to an ohm meter. 
Each of the seven steps should read around 50 Ω. 
The contacts on the ceramic attenuator elements can wear, they can be repaired with [[silver-bearing solder]]. 
Before reassembly, remove the two SMA housings to avoid damaging the contacts and remove the setscrews
that hold spring tension on the two ball bearings that align the shaft steps. 
Reinstall these parts after the housing has been screwed together.
 
==Pictures==
<gallery>
Tek1401a-front.jpg                                        | Front panel
Tek1401a-top.jpg                                          | Top PCB:  IF board
Tek1401a-bottom.jpg                                        | Bottom PCBs:gate/cal, power supply
Tek1401a-batterypcb.jpg                                    | PCB on the battery pack
Tek1401a-logamp.jpg                                        | [[Tektronix Cartoons | Cartoon in schematic diagram]]: "Log Amp"
Tek 1401A 324.jpg                                          | 1401A/324 combo from [[Media:Tekscope_1972_V4_N1_Jan_1972.pdf|Tekscope V.4 No.1 1972]]
John_Perry_323_and_140...jpg                              | 1401A/323 combo
John_Perry_323_and_1401A_side_view_with_custom_support.jpg | 1401A/323 combo side view with hand made acrylic support.
</gallery>


The input attenuator is a fairly custom part due to the IF gain pot
==Components==
that is concentric with the RF atten control.  Dissassembly of the
{{Parts|1401}}
attenuator is easy after removing it from the 1401A.  Note that the
{{Parts|1401A}}
IF gain pot is removed via two allen setscrews.  To bench test the
attenuator with DC, use a 50 ohm terminator


* [[List of manuals that are needed|Manual needed]]


[[Category:Spectrum Analyzers]]
[[Category:Spectrum analyzers]]
[[Category:Introduced in 1972]]
[[Category:Specifications needed]]
[[Category:Manual needed]]

Latest revision as of 12:53, 12 July 2024

Tektronix 1401
Spectrum analyzer front-end
Tektronix 1401A Spectrum Analyzer Module

Produced from 1971 to (?)

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix 1401 is a spectrum analyzer front-end covering the frequency range of 1 MHz to 500 MHz, designed to be used with an external oscilloscope, typically a 323, 324 or 326.

The 1401's case, like the 1501 TDR's, is the same size as the 323/324 case and designed to be stacked with the scope. It has internal rechargeable NiCd batteries and can also operate from AC power (using an unusual power cord) or external DC from 6 V to 16 V.

The external oscilloscope is assumed to have 10 horizontal divisions and 6 vertical divisions. Tektronix says it is compatible with any oscilloscope having 0.5 V/div horizontal deflection factor and 1.2 V full screen vertical deflection. Note that most full sized Tektronix oscilloscopes have 8 vertical divisions so use care when doing the initial setup and reading the scale in dBm. Also note the lack of a Z-axis output.

There appear to be two variants of the 1401A. Both the manual and catalog specify the resolution bandwidth selector as 1000, 100 and 3 kHz, but the later versions are 1000, 100, and 10 kHz. There is also a 1401A-1 which has a 75 Ω input impedance. The predecessor of the 1401A, the 1401, did not have the BNC "cal output" or "cal on" pushbutton switch.

Key Specifications

  • please add

Links

Documents Referencing 1401

Document Class Title Authors Year Links
TekWeek (partial) October 10, 1975.pdf Article Spectrum analyzers require high technology Thor Hallen Dave Friedley 1975

Internals

Operation is typical of a swept front end analyzer. The input is attenuated and low-pass filtered to 500 MHz. It is then mixed with the first LO, a varactor tuned oscillator that can sweep 695 MHz to 1195 MHz. The first mixer is a double balanced, four diode star. The signal is DC coupled to this mixer. The manual is not clear as to the damage level; with no attenuation they say that it is non linear above -30 dBm. The first IF output is 695 MHz, filtered to remove the other image. The first IF is mixed with a 720 MHz fixed 2nd LO to produce the 2nd IF at 25 MHz.

The 25 MHz 2nd IF leaves the shielded RF housing and enters the IF board via SMB connector P210. This 2nd IF is then amplified and mixed with a 30 MHz fixed 3rd LO to produce the 3rd IF of 5 MHz. This 3rd IF is then passed through the selectable RBW filters, detected and send to the video out.

Repair issues

See TekScope January 1972 pages 14-15 for some servicing notes that Tek wrote after the manual was published.

There is one tunnel diode found in the external trigger circuit. It is a 152-0402-00, a 2 mA, 25 pF tunnel diode in a DO-17 package. It is only used when the external trigger input is connected.

There is one Tektronix custom IC, 155-0042-02, listed as a Miller Integrator. It is used to generate the sweep ramp voltage. The package is a 10 pin metal can.

The input attenuator is a fairly custom part due to the IF gain pot that is concentric with the RF atten control. Dissassembly of the attenuator is easy after removing it from the 1401A. Note that the IF gain pot is removed via two allen setscrews. To bench test the attenuator with DC, use a 50 Ω terminator on SMA input and connect the output SMA to an ohm meter. Each of the seven steps should read around 50 Ω. The contacts on the ceramic attenuator elements can wear, they can be repaired with silver-bearing solder. Before reassembly, remove the two SMA housings to avoid damaging the contacts and remove the setscrews that hold spring tension on the two ball bearings that align the shaft steps. Reinstall these parts after the housing has been screwed together.

Pictures

Components

Some Parts Used in the 1401

Part Part Number(s) Class Description Used in
155-0028-00 155-0028-00 155-0028-01 155-0042-00 155-0042-01 155-0042-02 155-0042-03 Monolithic integrated circuit Miller integrator and delay pickoff 5030 R5030 5031 R5031 1401 1401A 1480 1481 1482 1485 26G1 26G2 26G3 314 335 432 434 4701 5B10N 5B12N 5B31 5B40 5B42 5S14N 7B52 7B53A 7B53N 7L12 7L13 7L14 7L18 7S14 AN/USM-281C RG501 Telequipment D63 Telequipment DM63
TD714 152-0402-00 Discrete component 2.2 mA, 25 pF tunnel diode 1401 1401A 283 R283 3B2 3B3 422 491 546 RM546 547 RM547 556 RM556 SPG11 SPG12


Some Parts Used in the 1401A

Part Part Number(s) Class Description Used in
155-0028-00 155-0028-00 155-0028-01 155-0042-00 155-0042-01 155-0042-02 155-0042-03 Monolithic integrated circuit Miller integrator and delay pickoff 5030 R5030 5031 R5031 1401 1401A 1480 1481 1482 1485 26G1 26G2 26G3 314 335 432 434 4701 5B10N 5B12N 5B31 5B40 5B42 5S14N 7B52 7B53A 7B53N 7L12 7L13 7L14 7L18 7S14 AN/USM-281C RG501 Telequipment D63 Telequipment DM63
TD714 152-0402-00 Discrete component 2.2 mA, 25 pF tunnel diode 1401 1401A 283 R283 3B2 3B3 422 491 546 RM546 547 RM547 556 RM556 SPG11 SPG12