Keithley 610B: Difference between revisions
Jump to navigation
Jump to search
(Created page with "{{Instrument Sidebar |manufacturer=Keithley |model=610B |class=Multimeter |series= |summary=Electrometer |image=Keithley 610B front.jpg |caption=Keithley 610B, frontal view |introduced=(?) |discontinued=(?) |designers= |manuals= * Keithley 610B Manual (OCR) }} The '''Keithley 610C State Electrometer''' can measure voltages (with a very high input resistance), very low currents, high resistances and charge. It used electrometer tubes a...") |
No edit summary |
||
Line 11: | Line 11: | ||
|designers= | |designers= | ||
|manuals= | |manuals= | ||
* [[Media:Keithley_610B_Manual.pdf|Keithley 610B Manual]] | * [[Media:Keithley_610B_Manual.pdf|Keithley 610B Manual]] | ||
}} | }} | ||
The '''Keithley 610C | The '''Keithley 610C Electrometer''' can measure voltages (with a very high input resistance), very low currents, high resistances and charge. There is a 610BR rack model as well. | ||
The 610B used electrometer tubes at the input and transistors in the gain stages. The predecessor, [[Keithley 610]], had used tubes in all stages, while the successor model, the [[Keithley 610C]], was fully solid-state. | |||
{{BeginSpecs}} | {{BeginSpecs}} | ||
Line 44: | Line 46: | ||
<gallery> | <gallery> | ||
Keithley 610B front.jpg | Keithley 610B front.jpg | ||
Keithley 610B rear.jpg | |||
</gallery> | </gallery> | ||
Revision as of 03:05, 1 September 2024
The Keithley 610C Electrometer can measure voltages (with a very high input resistance), very low currents, high resistances and charge. There is a 610BR rack model as well.
The 610B used electrometer tubes at the input and transistors in the gain stages. The predecessor, Keithley 610, had used tubes in all stages, while the successor model, the Keithley 610C, was fully solid-state.
Key Specifications
Voltage | 1 mV to 100 V full scale in 11 ranges (×1/×3), ±1% FS |
---|---|
Input resistance (volts mode) | >1014 Ω // 22 pF, or selectable 10 Ω to 1011 Ω in decade steps |
Noise | noise 10 μV with shorted input |
Zero offset drift | <2 mV in second hour after 1 hour warm-up |
Current | 10–14 A to 0.3 A full scale in 28 ranges (×1/×3), ±2% FS (≤3×10−12 A: ±4% FS) |
Resistance | 100 Ω to 1014 Ω full scale in 25 linear ranges (×1/×3), ±3% FS (≥3×109 Ω: ±5% FS) |
Charge | 10–13 C to 10–3 C (Ampere-seconds) full scale in 17 ranges (×1/×3), ±5% FS |
Display | Analog meter with mirror scale; positive, negative and center-zero display modes |
Outputs | Unity gain output, ±50ppm or 100 μV, max.1 mA; Recorder output, ±3 V full scale, polarity opposite to input, min. 300 Hz bandwidth; 3 kΩ source resistance |
Power | 105−125 V or 210−250 V, 50−1000 Hz, 10 W |
Accessories |
|
Internals
At the core of the 610B is a bootstrapped operational amplifier with a pair of 5886 electrometer tubes at the input, three differential transistor gain stages, and a 12AT7 in the output stage on a +250/–265 V supply.
Links
- Keithley 610B Electrometer @ radiomuseum.org
Pictures
Components
Some Parts Used in the Keithley 610B
Part | Part Number(s) | Class | Description | Used in |
---|---|---|---|---|
12AT7 | 154-0039-00 | Vacuum Tube (Dual Triode) | dual high-gain triode | 161 • 180 • 310 • 310A • 315 • 316 • 360 • 502 • 502A • 511A • 512 • 513 • 513D • 514 • 514AD • 514D • 516 • 524 • 529 • RM529 • 544 • 546 • 547 • 556 • 565 • 570 • 3A2 • 75 • 3A75 • 1M1 • A • B • C • G • H • K • L • ML • M • N • K • R • S • Z • Keithley 610B |
5886 | Vacuum Tube (Pentode) | electrometer pentode | Keithley 610 • Keithley 610B |