S-3
The Tektronix S-3 sampling head is a plug-in head for use with the 7000 series sampling plug-ins or the Type 3S2, 3S5, and 3S6 samplers.
The plug-in includes a permanently attached 1× probe with the sampling bridge enclosed in the probe. Optional 10× or 100× attenuator tips (010-0364-00, 010-0365-00) provide extended input voltage range; the optional VP-2 voltage pickoff converts it to a 50 Ω feed-through sampling head.
Key Specifications
Rise time | ≤350 ps (×10, ≤400 ps; ×100, ≤500 ps) |
---|---|
Input impedance | 100 kΩ // 2.3 pF (×10, 1 MΩ // 2 pF; ×100, 1 MΩ // 1.7 pF) |
Maximum signal | 2 Vp-p (×10, 20 Vp-p; ×100, 200 Vp-p) |
Maximum safe input | max. ±100 V, max. sine input 20 Vp-p (×10 or ×100, max. ±350 V, max. sine input 200 Vp-p) |
Features |
|
Links
Documents Referencing S-3
Document | Class | Title | Authors | Year | Links |
---|---|---|---|---|---|
Service Scope 53 Dec 1968.pdf | Article | Digital Systems Come Of Age | John Bowne | 1968 | 3T5 • 3T6 • 3S5 • 3S6 • S-1 • S-2 • S-3 • S-4 • 568 • 230 |
Service scope dec 1968 ocr.pdf | Article | Digital Systems Come of Age | John Bowne | 1968 | 3T5 • 3T6 • 3S5 • 3S6 • S-1 • S-2 • S-3 • S-4 • 568 • 230 • 240 • 241 • 250 |
Service Scope 52 Oct 1968.pdf | Article | The State of the Art in Sampling | Al Zimmerman | 1968 | S-1 • S-2 • S-3 • S-4 • S-50 • S-51 • 285 • 3S1 • 3S2 • 3S5 • 3S6 • 3T2 • 3T5 • 3T6 • 3T77A • 568 • 230 |
Differences between S-3 and S-3A
The S-3A sampling head has a 10× improvement in baseline shift due to cable flexing. All other specifications are identical to the ordinary S-3.
The major differences in design involves the strobe board in the box part of the probe eliminating the snap-off diode and adding a common mode choke in the S-3A , but some changes are also made to the circuits in the body of the probe.
Difference between the S-3 and the P6038
The S-3 probe contains a compensated attenuator before the sampling bridge. The P6038 does not. The S-3 probe has a JFET buffer between the holding cap and the probe cable. The P6038 does not. Both use the 152-0368-00 sampling bridge. The sampling pulse is generated by a snap-off diode (152-0368-00).
Repair of Sampling Diodes
Individuals have reported success in replacing the S-3 sampling diodes with Agilent matched pair Schottky diodes.
Pictures
S-3
-
S-3 probe sampling bridge
-
S-3 probe sampling bridge
-
S-3 schematic
-
010-0367-00 10X attenuator
S-3A
-
Front
-
S-3A probe
-
010-0364-00 ×10 attenuator for S-3A
-
S-3A probe internal
-
S-3A probe internal
-
S-3A strobe board
-
S-3A probe both sides
-
S-3A schematic
-
S-3A showing 280 ps risetime when probing a 50 Ω impedance with no probe attenuators.