Tektronix 067-0587-10
calibration fixture
067-0587-10

Compatible with 7000-series scopes

Produced from (?) to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 067-0587-10 is a calibration fixture / signal standardizer for 7000-series scopes. It has the same performance specifications as the 067-0587-02 but is modified for compatibility with the 7612D.

In the 067-0587-10, the signal coming from the PRETRIG OUT generates one trigger for each staircase signal when used with the TEST switch in the GAIN position. The 067-0587-02 generates a pre-trigger signal on the PRETRIG OUT connector while in any mode.

Key Specifications

     — GAIN mode
Staircase accuracy ± 0.3% in ± 3 Div, ± 0.5% in ± 4 Div, Linearity ± 0.2% of 10 Div
     — STEP RESP mode
Amplitude 2 to 10 Div
Position Top at least ±5 Div from center of screen
Rise time < 150 ps
Repetition rate 10 Hz to 1 MHz in decade steps, 0.1 %
Pretrig Out 0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable)
     — FREQ RESP mode
Input impedance 50 Ω
Input range 0.5 to 1 Vp-p
Flatness ± 3% from 3 MHz to 1 GHz

Pictures


Custom ICs used in the 067-0587-10

Page Model Part nos Description Designers Used in
155-0160-00 M136 155-0160-00 1.5 GHz trigger amplifier and source select Art Metz 7B10 7B15 067-0587-02 067-0587-10
155-0175-00 H476 155-0175-00 155-0175-05 broadband amplifier John Addis Wink Gross Carl Battjes 7904A 7912HB 7934 7104 R7103 7A29 7A29P 7F10 067-0587-02 067-0587-10 11A71 SCD1000 EG&G N-AM-173A
155-0182-00 H672 155-0182-00 Schottky diode shaper 067-0587-02 067-0587-10
155-0183-00 H672 155-0183-00 Schottky diode detector Art Metz 067-0587-02 067-0587-10