S-1

From TekWiki
(Redirected from S1)
Jump to navigation Jump to search
Tektronix S-1
Sampling Head
S-1 head

Compatible with 7000 and 3S series sampling heads

Produced from 1967 to 1989

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The S-1 sampling plug-in head provides a single GR-874 50 Ω input with a 350 ps rise-time.

The S-1 can be used with the 7000 series of sampling plug-ins, as well as the 3S series of sampling units.

Key Specifications

Rise time 350 ps
Bandwidth 1 GHz
Operating input voltage range −1 V to +1 V with ≤1 Vp-p signals
Maximum single sample step 500 mV
Maximum input voltage ±5 VDC or sine of 10 Vp-p up to 10 MHz
Input impedance 50 Ω ±1% (GR-874)
Noise < 2 mV
Features trigger pickoff for internal triggering

Links

Internals

The S-1 uses a two-diode sampling gate which has relatively high "kick-out" (leakage of the strobe signal out of the input).

The manual cautions,

Operating the sampling head without the input connector terminated by a 50 Ω resistor or coaxial cable will cause a vertical shift of the zero signal baseline by a few millivolts. This is caused by the strobe kickout signal being reflected from the open input connector, and arriving back at the sampling bridge while the bridge is conducting.

The sampling bridge strobe signal is generated by an avalanche transistor followed by a step recovery diode, called snap-off diode in Tek lingo.