S-3: Difference between revisions
No edit summary |
No edit summary |
||
Line 29: | Line 29: | ||
}} | }} | ||
{{EndSpecs}} | {{EndSpecs}} | ||
==Links== | |||
{{Documents|Link=S-3}} | |||
{{PatentLinks|S-3}} | |||
==Differences between S-3 and S-3A == | ==Differences between S-3 and S-3A == |
Revision as of 10:58, 19 June 2024
The Tektronix S-3 sampling head is a plug-in head for use with the 7000 series sampling plug-ins or the Type 3S2, 3S5, and 3S6 samplers.
The plug-in includes a permanently attached 1× probe with the sampling bridge enclosed in the probe. Optional 10× or 100× attenuator tips (010-0364-00, 010-0365-00) provide extended input voltage range; the optional VP-2 voltage pickoff converts it to a 50 Ω feed-through sampling head.
Key Specifications
Rise time | ≤350 ps (×10, ≤400 ps; ×100, ≤500 ps) |
---|---|
Input impedance | 100 kΩ // 2.3 pF (×10, 1 MΩ // 2 pF; ×100, 1 MΩ // 1.7 pF) |
Maximum signal | 2 Vp-p (×10, 20 Vp-p; ×100, 200 Vp-p) |
Maximum safe input | max. ±100 V, max. sine input 20 Vp-p (×10 or ×100, max. ±350 V, max. sine input 200 Vp-p) |
Features |
|
Links
Documents Referencing S-3
Document | Class | Title | Authors | Year | Links |
---|---|---|---|---|---|
Service Scope 53 Dec 1968.pdf | Article | Digital Systems Come Of Age | John Bowne | 1968 | 3T5 • 3T6 • 3S5 • 3S6 • S-1 • S-2 • S-3 • S-4 • 568 • 230 |
Service scope dec 1968 ocr.pdf | Article | Digital Systems Come of Age | John Bowne | 1968 | 3T5 • 3T6 • 3S5 • 3S6 • S-1 • S-2 • S-3 • S-4 • 568 • 230 • 240 • 241 • 250 |
Service Scope 52 Oct 1968.pdf | Article | The State of the Art in Sampling | Al Zimmerman | 1968 | S-1 • S-2 • S-3 • S-4 • S-50 • S-51 • 285 • 3S1 • 3S2 • 3S5 • 3S6 • 3T2 • 3T5 • 3T6 • 3T77A • 568 • 230 |
Differences between S-3 and S-3A
The S-3A sampling head has a 10× improvement in baseline shift due to cable flexing. All other specifications are identical to the ordinary S-3.
The major differences in design involves the gate board in the box part of the probe, but some changes are also made to the circuits in the body of the probe.
Difference between the S-3 and the P6038
The S-3 probe contains a compensated attenuator before the sampling bridge. The P6038 does not. The S-3 probe has a JFET buffer between the holding cap and the probe cable. The P6038 does not. Both use the 152-0368-00 sampling bridge. The sampling pulse is generated by a snap-off diode (152-0368-00).
Repair of Sampling Diodes
Individuals have reported success in replacing the S-3 sampling diodes with Agilent matched pair Schottky diodes.
Pictures
S-3
-
S-3 probe sampling bridge
-
S-3 probe sampling bridge
-
S-3 frontend circuit
S-3A
-
010-0364-00 ×10 attenuator for S-3A
-
Front
-
S-3A probe
-
S-3A probe internal
-
S-3A probe internal
-
S-3A strobe board
-
S-3A probe both sides
-
S-3A showing 280 ps risetime when probing a 50 Ω impedance with no probe attenuators.