P6042: Difference between revisions
No edit summary |
No edit summary |
||
(16 intermediate revisions by 3 users not shown) | |||
Line 1: | Line 1: | ||
{{Probe Sidebar | |||
The '''Tektronix P6042''' is a DC to 50 MHz current probe [[introduced in 1967]]. | |manufacturer=Tektronix | ||
|model=P6042 | |||
|summary=DC−50 MHz current probe | |||
|image=P6042-complete.jpg | |||
|caption=P6042 complete | |||
|for=50 Ω scope inputs | |||
|introduced=1967 | |||
|discontinued=(?) | |||
|designers=Cal Hongel; | |||
|manuals= | |||
* [[Media:070-0629-00.pdf|Tektronix P6042 Manual]] | |||
* [[Media:Tek p6042 calibration procedure dec 1967.pdf|Tektronix P6042 Calibration Procedure]] | |||
}} | |||
The '''Tektronix P6042''' is a DC to 50 MHz current probe [[introduced in 1967]]. It was designed by [[Cal Hongel]]. | |||
The sensor contains a current transformer for HF and a hall effect sensor for LF/DC components. | The sensor contains a current transformer for HF and a hall effect sensor for LF/DC components. | ||
The amplifier combines DC/LF and HF signals. | The amplifier combines DC/LF and HF signals. | ||
The [[AM503#Background|AM503]], introduced about a decade later, is a redesign of the P6042 with removable probes. | |||
{{BeginSpecs}} | {{BeginSpecs}} | ||
{{Spec | Bandwidth | DC to 50 MHz ( | {{Spec | Bandwidth | DC to 50 MHz (−3 dB) }} | ||
{{Spec | Rise time | 7 ns }} | {{Spec | Rise time | 7 ns }} | ||
{{Spec | Ranges | 1 mA/ | {{Spec | Ranges | 1 mA/div to 1 A/div (1–2–5) with scope in 50 mV/div }} | ||
{{Spec | Output | to 50 Ω scope input at 50 mV/ | {{Spec | Output | to 50 Ω scope input at 50 mV/div }} | ||
{{Spec | Maximum input | 10 A DC+peak AC in 5 mA to 1 A ranges, 0.5 A in 1 mA and 2 mA ranges }} | {{Spec | Maximum input | 10 A DC+peak AC in 5 mA to 1 A ranges, 0.5 A in 1 mA and 2 mA ranges }} | ||
{{Spec | Power | 90-136 or 180-272 V, 20 W}} | {{Spec | Power | 90-136 or 180-272 V, 20 W }} | ||
{{EndSpecs}} | {{EndSpecs}} | ||
== | ==Internals== | ||
Locking the probe requires pushing past the resistance felt when the core almost closes. This engages a mechanism that firmly presses both core halves together, and extinguishes the "Probe unlocked" indicator. | |||
The P6042's core halves are precision lapped and potted with epoxy. They have Mu-metal screening. The lower (fixed) core also contains the Hall element. | |||
Unfortunately, the P6042 probe is quite sensitive mechanically, and in many units offered for sale, the cores are damaged. | |||
The amplifier front-end is constructed using a 156-0014-00 differential amplifier integrated circuit, labelled "M18", which [[Media:Tektronix_Xref_sm.pdf|cross-references]] to a long-obsolete Teledyne "831-BE". [[Jim Williams]]' [https://www.analog.com/media/en/technical-documentation/application-notes/an61fa.pdf Applicaion Note 61] contains a replacement circuit using LT1228 opamps (page 19). | |||
==Prices== | |||
* $525 in 1967 (~$5,500 in 2022 Dollars) | |||
==Links== | ==Links== | ||
* [ | * [[Media:Service Scope 46 Oct 1967.pdf | Service Scope 46, October 1967]] | ||
* [https://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_other/P6042.htm Tek P6042 page @ amplifier.cd] | |||
* [http://www.barrytech.com/tektronix/probes/tekp6042.html Tek P6042 page @ barrytech.com] | * [http://www.barrytech.com/tektronix/probes/tekp6042.html Tek P6042 page @ barrytech.com] | ||
* [ | * [https://www.radiomuseum.org/r/tektronix_current_probe_dc_to_50_m.html Tek P6042 page @ radiomuseum.org] | ||
* [ | * [https://www.pa4tim.nl/?p=3135 P6042 repair by Fred, PA4TIM] | ||
* [ | * [https://www.eetimes.com/author.asp?section_id=36&doc_id=1330593 P6042 teardown @ EEtimes] | ||
{{Documents|Link=P6042}} | |||
{{PatentLinks|P6042}} | |||
==Pictures== | ==Pictures== | ||
<gallery> | <gallery> | ||
p6042-front.jpg | amplifier front | |||
p6042-complete.jpg | amplifier and sensor | |||
p6042-sensor.jpg | sensor detail | |||
p6042-back.jpg | rear view | |||
P6042.png | catalog image | |||
Tek p6042 amp internal1.jpg|P6042 amplifier internal view | |||
Tek p6042 amp internal2.jpg|P6042 amplifier internal view | |||
Tek p6042 amp internal3.jpg|P6042 amplifier internal view | |||
P6042 probe top.jpg | Top half core of probe | |||
P6042 probe bottom.jpg | Bottom of probe with half core and Hall sensor | |||
</gallery> | </gallery> | ||
[[Category:Oscilloscope current probes]] | [[Category:Oscilloscope current probes]] | ||
Latest revision as of 04:40, 13 July 2024
The Tektronix P6042 is a DC to 50 MHz current probe introduced in 1967. It was designed by Cal Hongel.
The sensor contains a current transformer for HF and a hall effect sensor for LF/DC components. The amplifier combines DC/LF and HF signals.
The AM503, introduced about a decade later, is a redesign of the P6042 with removable probes.
Key Specifications
Bandwidth | DC to 50 MHz (−3 dB) |
---|---|
Rise time | 7 ns |
Ranges | 1 mA/div to 1 A/div (1–2–5) with scope in 50 mV/div |
Output | to 50 Ω scope input at 50 mV/div |
Maximum input | 10 A DC+peak AC in 5 mA to 1 A ranges, 0.5 A in 1 mA and 2 mA ranges |
Power | 90-136 or 180-272 V, 20 W |
Internals
Locking the probe requires pushing past the resistance felt when the core almost closes. This engages a mechanism that firmly presses both core halves together, and extinguishes the "Probe unlocked" indicator.
The P6042's core halves are precision lapped and potted with epoxy. They have Mu-metal screening. The lower (fixed) core also contains the Hall element. Unfortunately, the P6042 probe is quite sensitive mechanically, and in many units offered for sale, the cores are damaged.
The amplifier front-end is constructed using a 156-0014-00 differential amplifier integrated circuit, labelled "M18", which cross-references to a long-obsolete Teledyne "831-BE". Jim Williams' Applicaion Note 61 contains a replacement circuit using LT1228 opamps (page 19).
Prices
- $525 in 1967 (~$5,500 in 2022 Dollars)
Links
- Service Scope 46, October 1967
- Tek P6042 page @ amplifier.cd
- Tek P6042 page @ barrytech.com
- Tek P6042 page @ radiomuseum.org
- P6042 repair by Fred, PA4TIM
- P6042 teardown @ EEtimes
Documents Referencing P6042
Document | Class | Title | Authors | Year | Links |
---|---|---|---|---|---|
Tekscope 1974 V6 N3.pdf | Article | Using your oscilloscope probe (Part 2) | Riley Stock | 1974 | P6201 • P6042 |
Patents that may apply to P6042
Page | Title | Inventors | Filing date | Grant date | Links |
---|---|---|---|---|---|
Patent US 3482163A | Magnetic signal measuring device including degaussing means | Bill Peek • Eldon Lee Hoven | 1967-05-24 | 1969-12-02 | P6042 • A6302 • A6302XL • A6303 • A6303XL • A6304XL • A6312 • AM503 • AM503A • AM503B • AM5030 • 11A16 |
Pictures
-
amplifier front
-
amplifier and sensor
-
sensor detail
-
rear view
-
catalog image
-
P6042 amplifier internal view
-
P6042 amplifier internal view
-
P6042 amplifier internal view
-
Top half core of probe
-
Bottom of probe with half core and Hall sensor