The '''Tektronix A6303''' aka '''P6303''' is a 15 MHz, 100 A current probe. | The '''Tektronix A6303''' aka '''P6303''' is a 15 MHz, 100 A current probe. | ||
It requires an [[AM503]], [[AM5030]] or [[11A16]] amplifier for operation. | It requires an [[AM503]]/A/B, [[AM5030]] or [[11A16]] amplifier for operation. | ||
The Tektronix [[067-0865-00]] | The Tektronix [[067-0865-00]] and [[067-0240-00]] are calibration fixtures for the P6303. | ||
{{BeginSpecs}} | {{BeginSpecs}} | ||
Line 34: | Line 34: | ||
{{Documents|Link=A6303}} | {{Documents|Link=A6303}} | ||
{{Documents|Link=P6303}} | {{Documents|Link=P6303}} | ||
{{PatentLinks|A6303}} | |||
==Pictures== | ==Pictures== |
Latest revision as of 04:45, 13 July 2024
The Tektronix A6303 aka P6303 is a 15 MHz, 100 A current probe.
It requires an AM503/A/B, AM5030 or 11A16 amplifier for operation. The Tektronix 067-0865-00 and 067-0240-00 are calibration fixtures for the P6303.
Key Specifications
Bandwidth | DC − 15 MHz |
---|---|
Rise time | ≤23 ns |
Input range | 100 A continuous, 500 A peak, ≤10 kA · μs (with 11A16: 52.5 A continuous)
Derating: 200 Ap-p up to 20 kHz, down to 20 Ap-p at 20 MHz (see curve) |
Sensitivity | 10 mA/Div to 50 A/Div, 1−2−5 (with AM503) |
Noise | ≤3 mA (@ 100 MHz amplifier BW, 10 mA range, DC) |
Links
- 120-1102-00 Current Transformer
- http://www.tek.com/datasheet/current-measurement-system-probes
- Tek A6303 @ barrytech.com
Documents Referencing A6303
- (no results)
Documents Referencing P6303
Document | Class | Title | Authors | Year | Links |
---|---|---|---|---|---|
Tekscope 1978 V10 N1.pdf | Article | New Products | 1978 | DF2 • P6303 • C-28 • 8001 • CT8100 • 4014-1 |
Patents that may apply to A6303
Page | Title | Inventors | Filing date | Grant date | Links |
---|---|---|---|---|---|
Patent US 3482163A | Magnetic signal measuring device including degaussing means | Bill Peek • Eldon Lee Hoven | 1967-05-24 | 1969-12-02 | P6042 • A6302 • A6302XL • A6303 • A6303XL • A6304XL • A6312 • AM503 • AM503A • AM503B • AM5030 • 11A16 |
Pictures
-
P6303 probe
-
Open jaws
-
A6303 maximum current vs frequency
-
A6303 insertion impedance vs frequency