The '''Tektronix A6303''' aka '''P6303''' is a 15 MHz, 100 A current probe.  
The '''Tektronix A6303''' aka '''P6303''' is a 15 MHz, 100 A current probe.  


It requires an [[AM503]], [[AM5030]] or [[11A16]] amplifier for operation.
It requires an [[AM503]]/A/B, [[AM5030]] or [[11A16]] amplifier for operation.
The Tektronix [[067-0865-00]] is a calibration fixture for the P6303.
The Tektronix [[067-0865-00]] and [[067-0240-00]] are calibration fixtures for the P6303.


{{BeginSpecs}}
{{BeginSpecs}}
Line 34: Line 34:
{{Documents|Link=A6303}}
{{Documents|Link=A6303}}
{{Documents|Link=P6303}}
{{Documents|Link=P6303}}
{{PatentLinks|A6303}}


==Pictures==
==Pictures==

Latest revision as of 04:45, 13 July 2024

Tektronix A6303 aka P6303
5 MHz, 100 A current probe
Tektronix A6303 aka P6303

Compatible with AM503

Produced from 1978 to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix A6303 aka P6303 is a 15 MHz, 100 A current probe.

It requires an AM503/A/B, AM5030 or 11A16 amplifier for operation. The Tektronix 067-0865-00 and 067-0240-00 are calibration fixtures for the P6303.

Key Specifications

Bandwidth DC − 15 MHz
Rise time ≤23 ns
Input range 100 A continuous, 500 A peak, ≤10 kA · μs (with 11A16: 52.5 A continuous)

Derating: 200 Ap-p up to 20 kHz, down to 20 Ap-p at 20 MHz (see curve)

Sensitivity 10 mA/Div to 50 A/Div, 1−2−5 (with AM503)
Noise ≤3 mA (@ 100 MHz amplifier BW, 10 mA range, DC)

Links

Documents Referencing A6303

(no results)

Documents Referencing P6303

Document Class Title Authors Year Links
Tekscope 1978 V10 N1.pdf Article New Products 1978

Patents that may apply to A6303

Page Title Inventors Filing date Grant date Links
Patent US 3482163A Magnetic signal measuring device including degaussing means Bill Peek Eldon Lee Hoven 1967-05-24 1969-12-02

Pictures