manufacturer=Tektronix |series=560-series scopes |type= 3B1 | | |manufacturer=Tektronix | ||
summary=10 MHz time base | | |series=560-series scopes | ||
image=Tek 3b1 front.jpeg | | |type= 3B1 | ||
caption=Tektronix 3B1 | |summary=10 MHz time base plug-in | ||
introduced=1962 | | |image=Tek 3b1 front.jpeg | ||
discontinued=(?) | | |caption=Tektronix 3B1 | ||
designers= | |introduced=1962 | ||
manuals= | |discontinued=(?) | ||
|designers= | |||
|manuals= | |||
* [[Media:070-344.pdf|Tektronix 3B1 Manual]] (PDF) | * [[Media:070-344.pdf|Tektronix 3B1 Manual]] (PDF) | ||
* [[Media:070-449.pdf|Tektronix 3B1 Performance Check Out]] | * [[Media:070-449.pdf|Tektronix 3B1 Performance Check Out]] |
Revision as of 23:20, 28 November 2023
The Tektronix 3B1 is a time-base plug-in introduced in 1962 for 560-series scopes. It is compatible with all the members of the 560 scope family other than the 560 itself.
There is a switch that enables 5 × horizontal magnification, which makes the fastest sweep 100 ns/Div. The 3B1 has a delayed sweep and a regular sweep.
Key Specifications
Trigger Frequency | DC to 10 MHz |
---|---|
Sweep Speeds | 1 Sec/div to 0.5 μs/div in 20 1-2-5 steps |
Sweep Types | Normal and Delayed |
Sweep Magnification | x5 |
Internals
The 3B1 uses vacuum tubes, bipolar junction transistors, and tunnel diodes.
The trigger input goes through the int/ext switch and then the AC/DC coupling switch, and then a 7895 cathode follower. The output of the cathode follower drives one base of a 2N2207 PNP differential amplifier which has a tunnel diode in the collector load. The tunnel diode generates pulses when it switches back and forth between its two stable voltages, which are capacitively coupled to a 2N2501 NPN common emitter amplifier.
The delay pickoff circuit is a 6DJ8 differential amplifier that compares the sweep ramp voltage with a knob-controlled DC voltage which determines the point where the delayed trigger starts. The DC voltage is made by a ZZ1000 80 V reference passed through a potentiometer-variable voltage divider.
When the sweep ramp exceeds the control voltage, a 2.2 mA tunnel diode in the plate load of the differential amplifier switches state.