Tek7104-ristetime-100hz.jpg | Rise time measurement. 150 ps risetime pulse from [[067-0587-02]] at 100 Hz repetition rate. Measured 240 ps (corrected for 150 ps fixture rise time: 190 ns = 1.8 GHz) | Tek7104-ristetime-100hz.jpg | Rise time measurement. 150 ps risetime pulse from [[067-0587-02]] at 100 Hz repetition rate. Measured 240 ps (corrected for 150 ps fixture rise time: 190 ns = 1.8 GHz) | ||
Tek7104-ristetime-1000hz.jpg | Rise time measurement as before, 1000 Hz repetition rate at same brightness settings. | Tek7104-ristetime-1000hz.jpg | Rise time measurement as before, 1000 Hz repetition rate at same brightness settings. | ||
Tek7104-200ps-singleshot.jpg | [[7104]] recording a single shot pulse (from [[067-0587-02]]) at 200 ps/Div. Camera: Nikon D7000, 50 mm f/1.4, ISO 3200, 1/2 s. CRT filter not removed. | |||
Tek7104-200ps-singleshot-sin1-1g.jpg | [[7104]] recording 1 GHz sine, single shot at 500 ps/Div. Camera: Nikon D7000, 50 mm f/1.4, ISO 3200, 1/2 s. CRT filter not removed. CRT amplification loss is evident around the center line. | Tek7104-200ps-singleshot-sin1-1g.jpg | [[7104]] recording 1 GHz sine, single shot at 500 ps/Div. Camera: Nikon D7000, 50 mm f/1.4, ISO 3200, 1/2 s. CRT filter not removed. CRT amplification loss is evident around the center line. | ||
Line 128: | Line 128: | ||
TD quick check example.jpg | Tunnel diode quick check example using the method from [[Media:Service Scope 49 Apr 1968.pdf | Service Scope 49]]. Right beam is zoomed with delayed timebase to show step speed. | TD quick check example.jpg | Tunnel diode quick check example using the method from [[Media:Service Scope 49 Apr 1968.pdf | Service Scope 49]]. Right beam is zoomed with delayed timebase to show step speed. | ||
7a29-rise.jpg | 7A29 rise time | |||
308-timing.jpg | [[308]] display, timing mode (binary counter example) | 308-timing.jpg | [[308]] display, timing mode (binary counter example) | ||
308-state.jpg | 308 display, state mode (binary counter example) | 308-state.jpg | 308 display, state mode (binary counter example) | ||
</gallery> | </gallery> |
Revision as of 10:20, 18 September 2017
Scopes
Plugins
Probes and other gear
-
P6451 probe internal, PCB front
-
P6451 probe internal, PCB back
Screen shots
-
Faster-than-light beam on 7104? Using a 067-0587-02 calibration fixture at maximum amplitude produces this trace spanning 1.2 Div horizontally at 200 ps/Div and 8 Div vertically. Trace length is (1.2² + 8²)½ × 8.5 mm = 68.8 mm, travelled in 1.2 × 200 ps = 240 ps. Apparent speed is therefore 68.8×10-3 / 240×10-12 m/s or 2.86×108 m/s. Very near miss ...
-
Rise time measurement. 150 ps risetime pulse from 067-0587-02 at 100 Hz repetition rate. Measured 240 ps (corrected for 150 ps fixture rise time: 190 ns = 1.8 GHz)
-
Rise time measurement as before, 1000 Hz repetition rate at same brightness settings.
-
7104 recording a single shot pulse (from 067-0587-02) at 200 ps/Div. Camera: Nikon D7000, 50 mm f/1.4, ISO 3200, 1/2 s. CRT filter not removed.
-
7104 recording 1 GHz sine, single shot at 500 ps/Div. Camera: Nikon D7000, 50 mm f/1.4, ISO 3200, 1/2 s. CRT filter not removed. CRT amplification loss is evident around the center line.
-
BC337 on 7CT1N
-
P6202 rise time
-
P6202 rise time
-
P6230 Rise time without ground lead: 9.2 * 20 ps = 184 ps (equiv. to 1.9 GHz bandwidth)
-
P6230 Rise time - Effect of using the "low inductance" ground lead - rise time 400 ps, bandwidth 875 MHz
-
P6230 Rise time - Using the 250 mm alligator ground lead - severe pulse distortion, rise time ~3 ns (115 MHz)!
-
A 7B92A (B098xxx) triggering a 974 MHz sine applied via 067-0587-02 calibration fixture
-
Classic readout board diagnostic display
-
Tektronix 067-0587-02, gain mode, in horizontal slot
-
Tektronix 067-0587-02, gain mode, in vertical slot
-
7B85 used for rise time measurement
-
7B85 used for rise time measurement
-
7B85 used for pulse width measurement
-
7B85 used for pulse period measurement
-
7B85 used for pulse period measurement
-
7S12 as a sampler displaying a 1 GHz signal (hi-res mode / storage)
-
7S12 as a sampler displaying a 1 GHz signal (hi-res mode / storage)
-
7S12 as a sampler displaying a 1 GHz signal (normal mode / non-storage)
-
7S12 displaying the incident pulse from an S-52 (nom. < 25 ps) through an S-6 head (nom. < 30 ps)
-
sub-millivolt triangle signal displayed through 7A22, BW=1 kHz
-
sub-millivolt triangle signal displayed through 7A22, BW=10 kHz
-
sub-millivolt triangle signal displayed through 7A22, BW=100 kHz
-
sub-millivolt triangle signal through 7A22, BW=1 MHz
-
Measuring diode reverse recovery
-
BAT41 diode reverse recovery, 20 mA/Div
-
Tunnel diode quick check example using the method from Service Scope 49. Right beam is zoomed with delayed timebase to show step speed.
-
7A29 rise time
-
308 display, timing mode (binary counter example)
-
308 display, state mode (binary counter example)