86: Difference between revisions
No edit summary |
No edit summary |
||
Line 11: | Line 11: | ||
|manuals= | |manuals= | ||
* [[Media:070-0364-01.pdf|Tektronix Type 86 Manual]] | * [[Media:070-0364-01.pdf|Tektronix Type 86 Manual]] | ||
* [ | * [[Media:Tek type 86 factory cal proc.pdf|Tektronix Type 86 Field Recalibration Procedure]] | ||
* [[Media:Tek type 86 fcp april 1967.pdf|Tektronix Type 86 Factory Calibration Procedure, April 1967]] | * [[Media:Tek type 86 fcp april 1967.pdf|Tektronix Type 86 Factory Calibration Procedure, April 1967]] | ||
}} | }} |
Latest revision as of 07:12, 22 December 2023
The Tektronix Type 86 is a single-input amplifier plug-in for 580-series scopes, introduced in 1963.
The P6008 voltage probe was designed for use with the Type 86.
Units with serial numbers before 1997 use a rotary input switch for the input coupling. Starting with serial number 1997, a lever switch is used.
Key Specifications
Input impedance | 1 MΩ // 15 pF |
---|---|
Bandwidth | 85 MHz at 100 mV/cm, 80 MHz at 10 mV/cm (in mainframe) |
Rise time | ×1 mode, 1.5 ns from 100 mV/Div to 20 V/Div; ×10 mode, 2.2 ns from 10 mV/Div to 2 V/Div |
Deflection | ×1 mode, 100 mV/Div to 50 V/Div; ×10 mode, 10 mV/Div to 5 V/Div; 1−2−5 |
Internals
The signal path in the Type 86 begins with a 7586 nuvistor as a cathode follower, feeding a differential transistor-based stage. In ×10 gain, this is followed by three more differential transistor gain stages. Finally, a distributed amplifier with three 6DJ8s drives the mainframe interface.
Like for the Type 82 plug-in, 585 non-A scopes with serial numbers before 2585 should have the back termination of the final distributed amplifier output modified to the later circuit (modification 040-275).
Prices
Year | 1965 | 1969 |
---|---|---|
Type 86 + P6008 | $350 | $385 |
In 2023 Dollars | $3,400 | $3,200 |
Pictures
-
-
-
-
-
Late serial number Type 86 with lever input coupling switch