7503: Difference between revisions

287 bytes added ,  13 November 2021
no edit summary
No edit summary
No edit summary
Line 39: Line 39:


==Internals==
==Internals==
The 7503 has an X-Y delay compensation network that is activated when the H plugin is an amplifier or a timebase in amplifier mode (see [[7000 Series plug-in interface]]). 
This network is switched in using a pair of [[148-0034-00]] miniature relays on the horizontal amplifier board.
Like other 75xx and 76xx mainframes and the 7504, the 7503 has a linear power supply, as opposed to the switchmode supplies used in the later 4-slot mainframes. In addition to the +5 V, ±15 V, ±50 V rails common to the 7000 series, internally (not on the plug-in slots) it has a +75 V output not found in later models, and a +150 V instead of the later +130 V.   
Like other 75xx and 76xx mainframes and the 7504, the 7503 has a linear power supply, as opposed to the switchmode supplies used in the later 4-slot mainframes. In addition to the +5 V, ±15 V, ±50 V rails common to the 7000 series, internally (not on the plug-in slots) it has a +75 V output not found in later models, and a +150 V instead of the later +130 V.