FG501A

Revision as of 23:42, 15 October 2024 by Peter (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

The Tektronix FG501A is a function generator plug-in for the TM500 system.

Tektronix FG501A
function generator
Tektronix FG501A

Produced from 1981 to 1993

Manuals
(All manuals in PDF format unless noted otherwise)

The FG501A replaced the FG501 (1972) in 1981.

Key Specifications

Frequency 0.002 Hz to 2 MHz in eight decade steps
Output Max. 30 Vp-p (into high impedance)
Attenuator 60 dB in 20 dB steps, 20 dB variable
Rise time 25 ns
Features symmetry variable from 5 to 95%
Signals sine, square, triangle, ramp, pulse
Features
  • voltage controlled frequency input (1000:1 ratio)
  • triggering
  • gated output with variable phase control

The TRIG OUT signal is available on the front panel via a BSM female connector.

The Phase control is effective in gated mode to control the starting phase of the generated waveform.

Differences to FG501

The FG501 non-A frequency maximum is 1 MHz, while the "A" model is specified at 2 MHz.

The FG501 features seven different functions – sine, triangle, square, positive- and negative-going ramps, pulse ≤50% duty, and pulse ≥50%. The duration of ramp and pulse waveforms is always equal to the half-cycle time of the sine, square, or triangle waveform frequency.

The FG501A model provides only three functions – sine, triangle, and square. However, it has an added variable symmetry control, effectively allowing the same functionality as the non-A model.

The FG501A also adds an output attenuator and increases the maximum open-circuit output voltage to 30 V from the FG501's 10/15 V.

Links

Documents Referencing FG501A

Document Class Title Authors Year Links
070-2088-04.pdf Book TM500 Series Rear Interface Data Book 1985

Patents that may apply to FG501A

Page Title Inventors Filing date Grant date Links
Patent US 3562464A Cam actuated switch having movable and fixed contacts on circuit board Howard Vollum Willem H Verhoef Tony Sprando 1968-10-07 1971-02-09

Rear Interface

  • Output: 28A (Common 27A)
  • Trigger Output: 27B (Common 28B)
  • Gate/Trigger Input: 24B (Common 25B)
  • VCF Input: 21B (Common 22B)

Pictures

Components

Some Parts Used in the FG501A

(no results)