AM503: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
(19 intermediate revisions by 5 users not shown)
Line 1: Line 1:
{{TM500 | type=AM503 | mfg=Tektronix | function=current probe amplifier | class=amplifier | image=Am503_front.jpg | introduced=1976 | discontinued=(?) | designers= |manuals=
{{TM500
'''Common'''
|type=AM503
|mfg=Tektronix
|function=current probe amplifier
|class=amplifier
|image=Am503_front.jpg
|introduced=1976
|discontinued=(?)
|designers=Dale Dorando
|manuals=
* [[Media:070-2052-01.pdf|AM503 Instruction Manual 070-2052-01]] (OCR, split schematics)
* [[Media:Tek AM503 diagrams.pdf|AM503 Diagrams]] (whole schematics)
* [[Media:AM50XX Declassification Documentrev4.pdf|AM50XX Declassification Document]]
* [[Media:AM50XX Declassification Documentrev4.pdf|AM50XX Declassification Document]]
'''AM503'''
* [[Media:070-2052-01.pdf|AM503 Instruction Manual 070-2052-01]] (OCR)
'''AM503A'''
* [[Media:070-8174-01.pdf|AM503A Instruction Manual 070-8174-01]]
'''AM503B/AM5030'''
* [[Media:070-8766-05.pdf|AM503B/AM5030 Instruction Manual 070-8766-05]]
* [[Media:070-8766-03.pdf|AM503B/AM5030 Instruction Manual 070-8766-03]]
* [[Media:070-8770-01.pdf|AM503B/AM5030 Reference 070-8770-01]]
* [[Media:070-9352-01.pdf|AM503B/5030, A6300 067-0271-00 Verification and Adjustment Kit Manual 070-9352-01]]
* [[Media:AM503B_AM5030_Calibration_Adapters.pdf|AM503B/AM5030 Calibration Adapters]]
'''AM503S'''
* [[Media:070-A245-00.pdf|AM503S User Manual - Japanese 070-A245-00]]
* [[Media:070-8170-00.pdf|AM503S User Manual 070-8170-00]]
'''Schematics'''
* [[Media:AM503A_Schematics.pdf|AM503A Schematics]]
}}
}}
Versions AM503A and AM503B have digital readout of deflection factor. The AM503S is an AM503A bundled with a [[TM502A]] mainframe, a [[016-0362-01|TM-500 toolbox]], and a [[A6302]] probe.
The AM503, [[AM503A]], [[AM503B]], [[AM5030]] and [[11A16]] use the same current probe input connector, an [[Amphenol 165-12 connector]]
It contains pins for hall element and transformer connections, lock detection, probe degaussing, and probe type encoding.


The current probe input is an [[Amphenol 165-12 connector]].  It contains pins for hall element and transformer connections, lock detection, probe degaussing, and probe type encoding.
The AM503 was designed by [[Dale Dorando]].


The [[AM503]], [[AM5030]] and [[11A16]] use the same connector.
Later facelift versions [[AM503A]] and [[AM503B]] have digital readout of deflection factor.


{{BeginSpecs}}
{{BeginSpecs}}
Line 47: Line 36:
* A6304XL: max. 500 A continuous, 700 A peak
* A6304XL: max. 500 A continuous, 700 A peak
}}
}}
{{Spec | Output | into 50 Ω scope input set to 10 mV/div }}
{{Spec | Output | Into 50 Ω Scope, Input Set to 10 mV/div }}
{{Spec | Power | 17 W}}
{{Spec | Power | 17 W }}
{{Spec | Compatible probes |
{{Spec | Compatible probes |
* [[A6302]] (20/100 A, DC-50 MHz), [[A6302XL]] (20/100 A, DC-17 MHz)
* [[A6302]] (20/100 A, DC-50 MHz)
* [[A6312]] (20/100 A, DC-100 MHz)
* [[A6312]] (20/100 A, DC-100 MHz)
* [[A6303]] (100/500 A, DC-15 MHz), [[A6303XL]] (100/500 A, DC-10 MHz)
* [[A6303]] (100/500 A, DC-15 MHz)
* [[A6304XL]] (500/700 A, DC-2 MHz, AM503B only)
* [[CT4]] with A6302 or A6312 (2000/20000 A)
* [[CT4]] with A6302, A6302XL or A6312 (2000/20000 A)
}}
}}
{{EndSpecs}}
{{EndSpecs}}
==Links==
{{Documents|Link=AM503}}
{{PatentLinks|AM503}}
===Rear interface===
===Rear interface===
The output can be detached from the front panel BNC socket which has a [[Peltola connector]] on the inside, and re-attached to a Peltola socket near the top of the interface connector, where it is terminated in 50 Ω and brought out on pin 28A (ground on 27A).
The output can be detached from the front panel BNC socket which has a [[Peltola connector]] on the inside, and re-attached to a Peltola socket near the top of the interface connector, where it is terminated in 50 Ω and brought out on pin 28A (ground on 27A).


==Background==
==Background==
Regarding the AM503, Dale Dorando says:
Regarding the AM503, designer [[Dale Dorando]] says:
 
<blockquote>
<blockquote>
The AM503 was a redesign of the [[P6042|6042]] current probe from 1967.   
The AM503 was a redesign of the [[P6042|6042]] current probe from 1967.   
It was initially going to be an oscilloscope plug-in.   
It was initially going to be an oscilloscope plug-in.   
About 80% through the project design a new line of products was started called TM500.
About 80% through the project design a new line of products was started called [[TM500]].
It was requested that we change the design to be a TM500 plug-in module instead.   
It was requested that we change the design to be a TM500 plug-in module instead.   
The AM503 front panel still looks like a scope’s plug-in.   
The AM503 front panel still looks like a scope’s plug-in.   
The AC/DC switch and 5 MHz bandwidth switch, for example, were left on the front panel from the oscilloscope design.
The AC/DC switch and 5 MHz bandwidth switch, for example, were left on the front panel from the oscilloscope design.
Line 73: Line 69:
The high frequency amplifier was differential to accommodate an oscilloscope’s differential input.   
The high frequency amplifier was differential to accommodate an oscilloscope’s differential input.   
The last change was the output stage that was added to convert the differential circuit to single ended, and then output that to a front panel BNC.  
The last change was the output stage that was added to convert the differential circuit to single ended, and then output that to a front panel BNC.  
<br />
 
The AM503 was also intended to work with future higher frequency probes, like the [[P6022]] with an added hall device for DC measurements.  
The AM503 was also intended to work with future higher frequency probes, like the [[P6022]] with an added hall device for DC measurements.  
The smaller P6022 had a 120 MHz bandwidth, whereas the [[P6302]] was only 50 MHz.   
The smaller P6022 had a 120 MHz bandwidth, whereas the [[P6302]] was only 50 MHz.   
The design engineer (actually only an R&D technician) moved to the TM500 engineering and evaluation group before this probe was created.  
The design engineer (actually only an R&amp;D technician) moved to the TM500 engineering and evaluation group before this probe was created.  
The AM503 was being developed in the Accessories division in Beaverton, since that’s where the other current probes were designed.   
The AM503 was being developed in the Accessories division in Beaverton, since that’s where the other current probes were designed.   
That was unusual as all the other plug-ins were being designed by the TM500 group on Walker Road, near Hillsboro.
That was unusual as all the other plug-ins were being designed by the TM500 group on Walker Road, near Hillsboro.
The AM503 was one of the first plug-ins for the TM500 line of products.   
The AM503 was one of the first plug-ins for the TM500 line of products.   
It required more power than the current plug-in designs so the transformer and supply were changed in the TM500 frames early on just for the AM503.
It required more power than the current plug-in designs, so the transformer and supply were changed in the TM500 frames early on just for the AM503.
<br />
 
There were some innovations that were incorporated in this design:   
There were some innovations that were incorporated in this design:   
The [[cam switch]] with replaceable laser trimmed attenuator resistors, a 1 GHz Gilbert cell op amp (actually developed by Howard Jones in 1963).
The [[cam switch]] with replaceable laser trimmed attenuator resistors, and a simple IC with a differential pair plus 1 GHz Gilbert cell (actually developed by Howard Jones in 1963).  
This IC, [[155-0078-xx]], was being developed by Tek’s IC department.
This IC, [[155-0078-xx]], was designed in Lab Scopes for the 485.  
(At the same time these IC designers were teaching the technology at the University of Portland as part of Tek’s excellent education and training program.)  
(At the same time designers were teaching amplifier technology internally and at the University of Portland as part of Tek’s excellent education and training program.)
The [[Peltola connector]], developed by Tek’s Ron Peltola, a very low cost connector that worked well, even for the highest frequencies.   
The [[Peltola connector]], developed by Tek’s Ron Peltola, a very low cost connector that worked well, even for the highest frequencies.   
A BNC was adapted to be used with the Peltola and is used on the front panel.
A BNC was adapted to be used with the Peltola and is used on the front panel.
<br />
 
A large section of the AM503 was basically duplicating the front end of an oscilloscope since that was the original design intent,  
A large section of the AM503 was basically duplicating the front end of an oscilloscope since that was the original design intent,  
so it was thought later that the differential amp section could be eliminated and the amplifier made much smaller and even become part of a standalone probe.   
so it was thought later that the differential amp section could be eliminated and the amplifier made much smaller and even become part of a standalone probe.   
A prototype was created (author’s senior project at the University of Portland) but shelved, as there was no longer an engineer in Accessories for current probes.   
A prototype was created (author’s senior project at the University of Portland) but shelved, as there was no longer an engineer in Accessories for current probes.   
Much later it gave rise to the TCP series of stand-alone current probes.
Much later it gave rise to the TCP series of stand-alone current probes.
<br />
 
The [[P6302]] and [[P6303]] probes utilized a Hall device that was manufactured in the clean room of the Accessories Manufacturing group.   
The [[P6302]] and [[P6303]] probes utilized a Hall device that was manufactured in the clean room of the Accessories Manufacturing group.   
It was deposited onto a bar of ferrite that was later assembled into a U shape with other ferrite and potted in a mu-metal can along with the transformer bobbins.   
It was deposited onto a bar of ferrite that was later assembled into a U shape with other ferrite and potted in a mu-metal can along with the transformer bobbins.   
Line 99: Line 95:
The cores were lapped and polished to a few Fresnel lines flatness to minimize the gap on the sliding ferrite.  
The cores were lapped and polished to a few Fresnel lines flatness to minimize the gap on the sliding ferrite.  
The L/R time constant affects the point where the Hall device and coils’ bandwidths crossover.
The L/R time constant affects the point where the Hall device and coils’ bandwidths crossover.
<br />
 
An interesting aspect of the design was trying to find a way to test and calibrate the peak current pulse on the larger P6303.   
An interesting aspect of the design was trying to find a way to test and calibrate the peak current pulse on the larger P6303.   
Luckily we had the tube lab.  We developed an argon filled thyratron that could discharge a 4 kV charge line into a 4 Ω load.   
Luckily we had the tube lab.  We developed an argon filled thyratron that could discharge a 4 kV charge line into a 4 Ω load.   
Line 107: Line 103:
The 4 kV supply was charged into a 4 Ω transmission line so a clean high current pulse would be generated.   
The 4 kV supply was charged into a 4 Ω transmission line so a clean high current pulse would be generated.   
This concept was taken from Tek’s [[109]] pulse generator.
This concept was taken from Tek’s [[109]] pulse generator.
<br />
 
A trigger circuit was designed to fire the grid.  It was based on the xenon flash circuit of the [[C-5|C5 camera]] flash, also in the Accessories group.
A trigger circuit was designed to fire the grid.  It was based on the xenon flash circuit of the [[C-5|C5 camera]] flash, also in the Accessories group.
The first prototype I made arced across the laser trim lines in the metal film resistor as they were cut perpendicular to the current flow.   
The first prototype I made arced across the laser trim lines in the metal film resistor as they were cut perpendicular to the current flow.   
Line 113: Line 109:
We changed the laser trim to be parallel to the current path along the outside edges.     
We changed the laser trim to be parallel to the current path along the outside edges.     
That’s what was used in manufacturing for calibrating the risetime of the P6303.
That’s what was used in manufacturing for calibrating the risetime of the P6303.
<br />
 
Also interesting, the P6303 required special potting epoxy developed by 3M.   
Also interesting, the P6303 required special potting epoxy developed by 3M.   
The epoxy used in other current probes put excessive stress on the larger ferrite that caused the inductance to drop to zero
The epoxy used in other current probes put excessive stress on the larger ferrite that caused the inductance to drop to zero
due to the magnetostriction property of ferrite. The epoxy could even also cause shear fractures in the ferrite.  
due to the magnetostriction property of ferrite. The epoxy could even also cause shear fractures in the ferrite.  
</blockquote>
</blockquote>
==Links==
* [[Media:Tekscope_1976_V8_N4_with_Supplement.pdf|TekScope Vol. 8 No. 4, Winter 1976 new products: P6302, AM503]]


==Pictures==
==Pictures==
====AM503====
<gallery>
<gallery>
Am503_front.jpg | AM503 front
Am503_front.jpg     | AM503 Front
Am503_left.jpg   | AM503 left side
Am503_left.jpg     | AM503 Left Side
Am503_right.jpg | AM503 right side
Am503_right.jpg     | AM503 Right Side
Tek am503.jpg   | AM503
Tek am503.jpg       | AM503
Tek am503 t1.jpg | AM503
Tek am503 t1.jpg   | AM503
Tek am503 t2.jpg | AM503 feeding a [[TDS744|TDS744A]]
Tek am503 t2.jpg   | AM503 Feeding a [[TDS744|TDS744A]]
Tek-sc504-am503.jpg | AM503 with [[SC504]] in a [[TM503]] mainframe displaying a 40 mA calibrator signal
AM503_Special_Adapter_Construction_Details.png | AM503 Special Adapter
AM503_Special_Adapter_Construction_Details.png | AM503 Special Adapter
</gallery>
====AM503A====
<gallery>
Tek am503a.jpg| AM503A
</gallery>
====AM503B====
<gallery>
Tek am503b.jpg        | AM503B
Tek am503b tm502a 1.jpg| Two AM503B in a [[TM502A]]
Tek am503b rear.jpg    | Rear view of two AM503B plug-ins
P1000607.JPG          | PCB component side
P1000609.JPG          | PCB soldering side
P1000611.JPG          | Ceramic daughter board with relays
P1000612.JPG          | 2nd daughter board
P1000613.JPG          | Tek specific IC
</gallery>
</gallery>


==See also==
==See also==
* [[AM5030]]
* [[AM5030]]
* http://www.tek.com/datasheet/current-measurement-system-probes
* https://www.tek.com/datasheet/current-measurement-system-probes
* Current Probe Simulator [[067-0802-99]]
* Current Probe Simulator [[067-0802-99]]


{{Custom ICs|AM503}}
==Components==
{{Parts|AM503}}

Revision as of 04:48, 13 July 2024

Tektronix AM503
current probe amplifier
Tektronix AM503

Produced from 1976 to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix AM503 is a current probe amplifier plug-in for the TM500 system.

The AM503, AM503A, AM503B, AM5030 and 11A16 use the same current probe input connector, an Amphenol 165-12 connector. It contains pins for hall element and transformer connections, lock detection, probe degaussing, and probe type encoding.

The AM503 was designed by Dale Dorando.

Later facelift versions AM503A and AM503B have digital readout of deflection factor.

Key Specifications

Bandwidth
Sensitivity
  • A6302, A6302XL, A6312: 1 mA/Div to 5 A/Div
  • A6303, A6303XL: 10 mA/Div to 50 A/Div
Range
  • A6302: max. 20 A continuous, 50 A peak
  • A6303: max. 100 A continuous, 500 A peak
  • A6304XL: max. 500 A continuous, 700 A peak
Output Into 50 Ω Scope, Input Set to 10 mV/div
Power 17 W
Compatible probes
  • A6302 (20/100 A, DC-50 MHz)
  • A6312 (20/100 A, DC-100 MHz)
  • A6303 (100/500 A, DC-15 MHz)
  • CT4 with A6302 or A6312 (2000/20000 A)

Links

Documents Referencing AM503

Document Class Title Authors Year Links
Tekscope 1976 V8 N4 with Supplement.pdf Article New Products 1976
070-2088-04.pdf Book TM500 Series Rear Interface Data Book 1985

Patents that may apply to AM503

Page Title Inventors Filing date Grant date Links
Patent US 3482163A Magnetic signal measuring device including degaussing means Bill Peek Eldon Lee Hoven 1967-05-24 1969-12-02
Patent US 3562464A Cam actuated switch having movable and fixed contacts on circuit board Howard Vollum Willem H Verhoef Tony Sprando 1968-10-07 1971-02-09
Patent US 3900711A Electrical contactor assembly for high frequency applications Ken Holland 1974-02-04 1975-08-19

Rear interface

The output can be detached from the front panel BNC socket which has a Peltola connector on the inside, and re-attached to a Peltola socket near the top of the interface connector, where it is terminated in 50 Ω and brought out on pin 28A (ground on 27A).

Background

Regarding the AM503, designer Dale Dorando says:

The AM503 was a redesign of the 6042 current probe from 1967.

It was initially going to be an oscilloscope plug-in. About 80% through the project design a new line of products was started called TM500. It was requested that we change the design to be a TM500 plug-in module instead.

The AM503 front panel still looks like a scope’s plug-in. The AC/DC switch and 5 MHz bandwidth switch, for example, were left on the front panel from the oscilloscope design. There are still some remnants of the oscilloscope circuits, such as the cam switch that interfaced to a circuit board that sets the scope’s readout in amps/div. The square pin header connector on the back of the circuit board was for the readout, but not used. The high frequency amplifier was differential to accommodate an oscilloscope’s differential input. The last change was the output stage that was added to convert the differential circuit to single ended, and then output that to a front panel BNC.

The AM503 was also intended to work with future higher frequency probes, like the P6022 with an added hall device for DC measurements. The smaller P6022 had a 120 MHz bandwidth, whereas the P6302 was only 50 MHz. The design engineer (actually only an R&D technician) moved to the TM500 engineering and evaluation group before this probe was created. The AM503 was being developed in the Accessories division in Beaverton, since that’s where the other current probes were designed. That was unusual as all the other plug-ins were being designed by the TM500 group on Walker Road, near Hillsboro. The AM503 was one of the first plug-ins for the TM500 line of products. It required more power than the current plug-in designs, so the transformer and supply were changed in the TM500 frames early on just for the AM503.

There were some innovations that were incorporated in this design: The cam switch with replaceable laser trimmed attenuator resistors, and a simple IC with a differential pair plus 1 GHz Gilbert cell (actually developed by Howard Jones in 1963). This IC, 155-0078-xx, was designed in Lab Scopes for the 485. (At the same time designers were teaching amplifier technology internally and at the University of Portland as part of Tek’s excellent education and training program.) The Peltola connector, developed by Tek’s Ron Peltola, a very low cost connector that worked well, even for the highest frequencies. A BNC was adapted to be used with the Peltola and is used on the front panel.

A large section of the AM503 was basically duplicating the front end of an oscilloscope since that was the original design intent, so it was thought later that the differential amp section could be eliminated and the amplifier made much smaller and even become part of a standalone probe. A prototype was created (author’s senior project at the University of Portland) but shelved, as there was no longer an engineer in Accessories for current probes. Much later it gave rise to the TCP series of stand-alone current probes.

The P6302 and P6303 probes utilized a Hall device that was manufactured in the clean room of the Accessories Manufacturing group. It was deposited onto a bar of ferrite that was later assembled into a U shape with other ferrite and potted in a mu-metal can along with the transformer bobbins. The Hall device used for the DC measurements used a vacuum deposition process with indium antimonide. The cores were lapped and polished to a few Fresnel lines flatness to minimize the gap on the sliding ferrite. The L/R time constant affects the point where the Hall device and coils’ bandwidths crossover.

An interesting aspect of the design was trying to find a way to test and calibrate the peak current pulse on the larger P6303. Luckily we had the tube lab. We developed an argon filled thyratron that could discharge a 4 kV charge line into a 4 Ω load. The load resistor was designed and built by Tektronix. A large rectangular ceramic plate was coated with a metal film resistor. It had a voltage divider tap to allow for a safer measurement point. It was laser trimmed for accuracy. The current probe would measure the current to ground through this resistor, so the voltage was near zero for the user. The 4 kV supply was charged into a 4 Ω transmission line so a clean high current pulse would be generated. This concept was taken from Tek’s 109 pulse generator.

A trigger circuit was designed to fire the grid. It was based on the xenon flash circuit of the C5 camera flash, also in the Accessories group. The first prototype I made arced across the laser trim lines in the metal film resistor as they were cut perpendicular to the current flow. This caused high voltage gradients across the film and thus the arcing across the cuts. We changed the laser trim to be parallel to the current path along the outside edges. That’s what was used in manufacturing for calibrating the risetime of the P6303.

Also interesting, the P6303 required special potting epoxy developed by 3M. The epoxy used in other current probes put excessive stress on the larger ferrite that caused the inductance to drop to zero due to the magnetostriction property of ferrite. The epoxy could even also cause shear fractures in the ferrite.

Pictures

See also

Components

Some Parts Used in the AM503

Part Part Number(s) Class Description Used in
155-0078-00 155-0078-xx 155-0273-00 155-0274-00 Monolithic integrated circuit broadband amplifier 464 465 466 468 475 475A 475M 485 7834 7844 7854 7904 R7903 R7912 7912AD 7912HB 7104 7A16A 7A16P 7A24 7A26 7A42 067-0587-01 067-0680-00 AM503 PG502 PG508 DC510 DC5010 FG5010